IBC FLOW PUMP
MDR report key: 347700
·
Received August 3, 2001
Report
- Report Number
- 347700
- Event Type
- Injury
- Date Received
- August 3, 2001
- Date of Event
- August 3, 2001
- Manufacturer
- INTERNATIONAL BIOPHYSICS CORP.
- Product Code
- KFM
- Report Source
- User Facility report
- Reporter Location
- CA, US
- Reporter Occupation
- RISK MANAGER
Narratives
Description of Event or Problem · 1
APPROX 20 MINUTES INTO A ONE HOUR CABG, THE IBC CONE BEGAN TO SHAKE & MAKE NOISE. REMOVING THE CONE FROM THE DRIVE TO RECOUPLE THE MAGNETS DID NOT IMPROVE THE PROBLEM OF THE NOISE. FLOW WAS MAINTAINED AT A CARDIAC INDEX OF 26-2.8 UNTIL THE CONE FAILED. THE A&V LINES WERE CLAMPED, AND THE HEART WAS MANIPULATED BY THE FIELD AND VENTILATION MADE BY ANESTHESIA. A NEW CONE WAS CUT INTO PLACE, PRIMED, AND THE PT WAS BACK ON BYPASS IN LESS THAN 2 MINUTES. ALL MANUFACTURERS OF DIPOSABLES &/OR THE PUMP CONSOLE HAVE BEEN NOTIFIED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 35735 | IBC FLOW PUMP | CENTRIFUGAL PUMP | KFM | INTERNATIONAL BIOPHYSICS CORP. | 6000 | 135101 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 75 YR | Required Intervention |