FDA Adverse Event Other Summary report: N

ARTAS

MDR report key: 3472081 · Received October 30, 2013

Report

Report Number
3006515340-2013-00001
Event Type
Other
Date Received
October 30, 2013
Date of Event
October 8, 2013
Report Date
October 29, 2013
Manufacturer
RESTORATION ROBOTICS
Product Code
ONA
PMA / PMN Number
K123548
Removal / Correction Number
3006515340-10/17/13-001C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

ADD'L MODEL NUMBERS: (B)(4). DISCOLORATION DUE TO INSUFFICIENT CLEANING OF STAINLESS STEEL PINS DURING PASSIVATION PROCESS. THERE HAVE BEEN NO ADVERSE EVENT ASSOCIATED WITH THIS ISSUE. USERS NOTIFICATION INITIATED 8 OCT 2013. CORRECTION 3006515340-10/17/13-001C, REPORTED TO (B)(4) FIELD OFFICE (B)(4) 2013.

Description of Event or Problem · 1

DISCOLORATION OBSERVED ON PINS USED IN DISPOSABLE SKIN TENSIONER. TO DATE NO SERIOUS INJURY HAS BEEN REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
556283 ARTAS ARTAS SYSTEM ONA RESTORATION ROBOTICS FP-45616 NA

Patients

Seq Age Sex Outcome Treatment
1 NA