FDA Adverse Event Injury Summary report: N

MEDICAL ILLUMINATION INT'L, INC.

MDR report key: 347056 · Received August 2, 2001

Report

Report Number
2028295-2001-00002
Event Type
Injury
Date Received
August 2, 2001
Date of Event
June 5, 2001
Report Date
July 13, 2001
Manufacturer
MEDICAL ILLUMINATION INC.
Product Code
KZF
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
PHYSICIAN

Narratives

Description of Event or Problem · 1

THE DOG WAS PREPPED FOR SURGERY WHEN THE LIGHT WAS TURNED ON. THE BULB BURNED OUT WHEN IT WAS TURNED ON AND DR WENT TO REPLACE THE BULB, INSTEAD OF REMOVING THE HANDLE BASE SCREWS TO GAIN ACCESS TO THE BULB, DR REMOVED THE SPRINGS THAT HOLDS THE FILTER IN PLACE. THEY THEN HAD DIFFICULTY RE-ASSMEBLING THE LIGHT, AND SINCE THE DOG WAS ANESTHETIZED, DECIDED TO LEAVE THE FILTER OFF AND PROCEED WITH THE SURGERY. THE DOG RECEIVED A THERMAL BURN MEASURING APPROXIMATELY 6"X9". THE DOG HAS BEEN TREATED FOR THE BURN BY CHANGING THE DRESSING DAILY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
35013 MEDICAL ILLUMINATION INT'L, INC. LIGHT, MEDICAL EXAMINATION KZF MEDICAL ILLUMINATION INC. 011314-6 *

Patients

Seq Age Sex Outcome Treatment
1 10 YR