FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 3468204 · Received November 13, 2013

Report

Report Number
9611451-2013-00908
Event Type
Malfunction
Date Received
November 13, 2013
Date of Event
October 8, 2013
Report Date
October 17, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZE
PMA / PMN Number
K983112
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT WAS RETURNED TO FPH IN (B)(4) FOR INSPECTION. IT WAS PRESSURE TESTED, VISUALLY INSPECTED, AND IMMERSED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: THE PRESSURE TEST RESULT WAS OUT OF SPECIFICATION DUE TO LEAKAGE COMING FROM THE PATIENT END CONNECTOR. THIS WAS CONFIRMED WHEN THE SUBJECT BREATHING CIRCUIT WAS IMMERSED IN THE WATER BATH. VISUAL INSPECTION REVEALED THAT THERE WAS NOT ENOUGH GLUE PRESENT IN THE EVAQUA EXPIRATORY LIMB CONNECTOR, CAUSING THE REPORTED LEAK. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 130719. CONCLUSION: ALL BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED BEFORE RELEASING FOR DISTRIBUTION, AND THOSE THAT FAIL ARE REJECTED. THE OBSERVED LEAK WAS CAUSED BY INSUFFICIENT GLUE BEING INJECTED INTO THE EVAQUA LIMB CONNECTOR SUCH THAT IT DID NOT FORM A PERMANENT SEAL DURING THE ASSEMBLY PROCESS. THE USER INSTRUCTIONS SUPPLIED WITH ADULT EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS." (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT AN RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT FAILED THE VENTILATOR LEAK TEST. THIS WAS OBSERVED BEFORE USE ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
587396 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LIMITED RT340 130719

Patients

Seq Age Sex Outcome Treatment
1