FDA Adverse Event
Malfunction
Summary report: N
HITACHI 911
MDR report key: 34665
·
Received July 29, 1996
Report
- Report Number
- MW1009586
- Event Type
- Malfunction
- Date Received
- July 29, 1996
- Date of Event
- April 30, 1996
- Report Date
- July 26, 1996
- Manufacturer
- BOEHRINGER-MANNHEIM CORP.
- Product Code
- JJC
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
SV-12 INPUT NIPPLE LEAKED WASH REAGENT (NAOH). DUE TO PLACEMENT OF THIS CONNECTION, NAOH CONTACTED HOUSING OF VACUUM PUMP DESTROYING HOUSING, RESULTING IN ANALYZER FAILURE. RAPID IDENTIFICATION OF PROBLEM BY TECHNOLOGIST PREVENTED POSSIBLE FIRE. THE DESIGN OF THE INSTRUMENT PRESENTS A SIGNIFICANT SAFETY RISK.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | HITACHI 911 | AUTOMATED CHEMISTRY ANALYZER | JJC | BOEHRINGER-MANNHEIM CORP. | 911 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO | Other |