FDA Adverse Event Malfunction Summary report: N

HITACHI 911

MDR report key: 34665 · Received July 29, 1996

Report

Report Number
MW1009586
Event Type
Malfunction
Date Received
July 29, 1996
Date of Event
April 30, 1996
Report Date
July 26, 1996
Manufacturer
BOEHRINGER-MANNHEIM CORP.
Product Code
JJC
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

SV-12 INPUT NIPPLE LEAKED WASH REAGENT (NAOH). DUE TO PLACEMENT OF THIS CONNECTION, NAOH CONTACTED HOUSING OF VACUUM PUMP DESTROYING HOUSING, RESULTING IN ANALYZER FAILURE. RAPID IDENTIFICATION OF PROBLEM BY TECHNOLOGIST PREVENTED POSSIBLE FIRE. THE DESIGN OF THE INSTRUMENT PRESENTS A SIGNIFICANT SAFETY RISK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 HITACHI 911 AUTOMATED CHEMISTRY ANALYZER JJC BOEHRINGER-MANNHEIM CORP. 911

Patients

Seq Age Sex Outcome Treatment
1 NO INFO Other