ADULT DUAL HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00905
- Event Type
- Malfunction
- Date Received
- November 12, 2013
- Date of Event
- October 5, 2013
- Report Date
- October 17, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT104 ADULT DUAL HEATED BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT THAT IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE COMPLAINT RT104 ADULT BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR INSPECTION. THE ELECTRICAL RESISTANCE OF THE INSPIRATORY AND THE EXPIRATORY HEATERWIRES WERE TESTED USING A CALIBRATED MULTIMETER. CONTINUITY TEST AND VISUAL INSPECTION WERE ALSO CONDUCTED TO CHECK THE ELECTRICAL CONNECTION BETWEEN THE HEATERWIRES AND THE HEATERWIRE PINS. RESULTS: ELECTRICAL RESISTANCE TEST SHOWED THE EXPIRATORY HEATERWIRE WAS WITHIN SPECIFICATION; HOWEVER, THE INSPIRATORY HEATERWIRE WAS OPEN CIRCUIT. CONTINUITY TEST AND VISUAL INSPECTION REVEALED THAT THE OPEN CIRCUIT IN THE INSPIRATORY HEATERWIRE WAS LOCATED AT THE CONNECTION BETWEEN THE HEATERWIRE AND THE LEFT HEATERWIRE PIN INSIDE THE OVERMOULDED PLUG. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 130604. CONCLUSION: ELECTRICAL OPEN CIRCUITS IN HEATERWIRES CAN BE ASSOCIATED WITH INSUFFICIENT CONNECTION OF THE HEATERWIRE PIN DURING PRODUCTION, SUCH THAT IT IS UNABLE TO PROVIDE CONTINUITY FOR THE FULL PERIOD OF USE. ALL RT104 BREATHING CIRCUITS ARE RESISTANCE AND CONTINUITY TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE SUBJECT INSPIRATORY HEATERWIRE BECAME OPEN CIRCUIT AFTER RELEASED FOR DISTRIBUTION, POSSIBLY AS A RESULT OF INSUFFICIENT CONNECTION. (B)(4).
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT A "HEATER WIRE DISCONNECT" ALARM SOUNDED ON THE MR850 RESPIRATORY HUMIDIFIER. THIS WAS OBSERVED WHEN THEY STARTED USING AN RT104 ADULT DUAL HEATED BREATHING CIRCUIT, WHICH WAS CONNECTED TO THE MR850 HUMIDIFIER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 581989 | ADULT DUAL HEATED BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | RT104 | 1306040304 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | MR850 RESPIRATORY HUMIDIFIER |