FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED BREATHING CIRCUIT

MDR report key: 3465470 · Received November 12, 2013

Report

Report Number
9611451-2013-00905
Event Type
Malfunction
Date Received
November 12, 2013
Date of Event
October 5, 2013
Report Date
October 17, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT104 ADULT DUAL HEATED BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT THAT IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE COMPLAINT RT104 ADULT BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR INSPECTION. THE ELECTRICAL RESISTANCE OF THE INSPIRATORY AND THE EXPIRATORY HEATERWIRES WERE TESTED USING A CALIBRATED MULTIMETER. CONTINUITY TEST AND VISUAL INSPECTION WERE ALSO CONDUCTED TO CHECK THE ELECTRICAL CONNECTION BETWEEN THE HEATERWIRES AND THE HEATERWIRE PINS. RESULTS: ELECTRICAL RESISTANCE TEST SHOWED THE EXPIRATORY HEATERWIRE WAS WITHIN SPECIFICATION; HOWEVER, THE INSPIRATORY HEATERWIRE WAS OPEN CIRCUIT. CONTINUITY TEST AND VISUAL INSPECTION REVEALED THAT THE OPEN CIRCUIT IN THE INSPIRATORY HEATERWIRE WAS LOCATED AT THE CONNECTION BETWEEN THE HEATERWIRE AND THE LEFT HEATERWIRE PIN INSIDE THE OVERMOULDED PLUG. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 130604. CONCLUSION: ELECTRICAL OPEN CIRCUITS IN HEATERWIRES CAN BE ASSOCIATED WITH INSUFFICIENT CONNECTION OF THE HEATERWIRE PIN DURING PRODUCTION, SUCH THAT IT IS UNABLE TO PROVIDE CONTINUITY FOR THE FULL PERIOD OF USE. ALL RT104 BREATHING CIRCUITS ARE RESISTANCE AND CONTINUITY TESTED DURING PRODUCTION, AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT THE SUBJECT INSPIRATORY HEATERWIRE BECAME OPEN CIRCUIT AFTER RELEASED FOR DISTRIBUTION, POSSIBLY AS A RESULT OF INSUFFICIENT CONNECTION. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE (FPH) FIELD REPRESENTATIVE THAT A "HEATER WIRE DISCONNECT" ALARM SOUNDED ON THE MR850 RESPIRATORY HUMIDIFIER. THIS WAS OBSERVED WHEN THEY STARTED USING AN RT104 ADULT DUAL HEATED BREATHING CIRCUIT, WHICH WAS CONNECTED TO THE MR850 HUMIDIFIER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
581989 ADULT DUAL HEATED BREATHING CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED RT104 1306040304

Patients

Seq Age Sex Outcome Treatment
1 MR850 RESPIRATORY HUMIDIFIER