FDA Adverse Event Malfunction Summary report: N

ANTI-JKB BIOCLONE ANTISERA

MDR report key: 3457411 · Received November 8, 2013

Report

Report Number
2250051-2013-00389
Event Type
Malfunction
Date Received
November 8, 2013
Date of Event
September 18, 2013
Report Date
November 8, 2013
Manufacturer
ORTHO CLINICAL DIAGNOSTICS
Product Code
KSZ
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

OCD PERFORMED RETAIN TESTING, BATCH REVIEW AND COMPLAINT REVIEW BY LOT. ALL RESULTS WERE SATISFACTORY. TESTING OF RETURNED VIALS OF ANTISERA LOT #JBB423A WAS ALSO PERFORMED. SATISFACTORY RESULTS WERE OBTAINED. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED FALSE NEGATIVE RESULTS WITH DONOR # (B)(4) AND JKB ANTIGEN TYPING PERFORMED WITH THE ORTHO ANTI-JKB LOT# JBB423A. CUSTOMER REPORTS THE INITIAL TESTING WAS PERFORMED ON (B)(6) 2013 WHERE DONOR # (B)(6) WAS TESTED AGAINST THE ORTHO ANTI-JKB LOT# JBB423A AND WAS REPORTED AS BEING 1+ POSITIVE FOR THE JKB ANTIGEN. QUALITY CONTROLS WERE CONFIRMED TO BE ACCEPTABLE WITH IMMUCOR HETEROZYGOUS DONORS ON THE DAY OF USE. CUSTOMER STATED THAT THE LISTED DONOR WAS SELECTED TO BE FROZEN BASED ON ADDITIONAL ANTIGEN TYPING. CUSTOMER'S SOP CALLS FOR ALL ANTIGEN TYPING TO BE REPEATED BEFORE THE UNITS WILL BE FROZEN. CUSTOMER REPORTED THIS REPEAT JKB TYPING WAS PERFORMED ON (B)(6) 2013 AND THE JKB TYPING WERE INCONSISTENT WITH THE ORIGINAL RESULTS. DONOR # (B)(6) WAS NOW YIELDING NO REACTIVITY AGAINST THE SAME LOT OF ANTI-JKB. CUSTOMER REPORTED MAKING MULTIPLE ATTEMPTS WITH TESTING THE DONOR AND RESULTS WOULD BE INCONSISTENTLY POSITIVE AND NEGATIVE. QUALITY CONTROL WAS CONFIRMED TO BE ACCEPTABLE FOR THE DAY OF USE. ALL REAGENTS HAD BEEN STORED ACCORDING TO THEIR IFU AND HAD NORMAL APPEARANCE. CUSTOMER CONFIRMED THAT THE CORRECT PROCEDURE FOR TRADITIONAL TUBE METHOD WAS BEING FOLLOWED WITH A 30MIN ROOM TEMP INCUBATION. CUSTOMER INDICATED THAT THE UNIT WOULD NOT BE USED FOR ANY PATIENT KNOW TO HAVE AN ANTI-JKB. CUSTOMER REQUESTED THE REPLACEMENT OF BOTH VIALS OF THE JKB ANTISERA FOR AN ALTERNATIVE LOT. REPLACEMENT ANTI-JKB LOT # JBB424A WAS SUPPLIED TO THE CUSTOMER. CUSTOMER INDICATED THEIR CONFIDENCE THAT THE DONOR IS A JKB VARIANT AND THIS IS THE CAUSE OF THE INCONSTANCIES. CUSTOMER REPORTED SENDING THE DONOR SEGMENT OUT FOR MOLECULAR JKB STUDIES. NO ADDITIONAL TESTING OF THE DONOR BY OCD WAS REQUIRED. CUSTOMER AGREED TO RETURN VIALS OF ANTI-JKB ANTISERA LOT# JBB423A FOR TESTING BY OCD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
579816 ANTI-JKB BIOCLONE ANTISERA RED BLOOD CELL REAGENT KSZ ORTHO CLINICAL DIAGNOSTICS JBB423A

Patients

Seq Age Sex Outcome Treatment
1