FDA Adverse Event Malfunction Summary report: N

LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT

MDR report key: 3457088 · Received November 7, 2013

Report

Report Number
2955842-2013-05089
Event Type
Malfunction
Date Received
November 7, 2013
Date of Event
July 24, 2013
Report Date
October 10, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. FAILURE ANALYSIS FOUND THE INSTRUMENT PASSED AN INTUITIVE MOTION TEST ON AN (B)(4) SYSTEM AND NO CLICK SOUND WAS HEARD. AN ADDITIONAL FINDING WAS VARIOUS SCRATCHES ON THE DISTAL END OF THE MAIN TUBE SHOWING LIGHT MATERIAL REMOVAL AND A ROUGH SURFACE FINISH. THE SCRATCHES WERE .031 - .127 IN LENGTH AND NOT AXIALLY ALIGNED WITH THE TUBE. FAILURE ANALYSIS CONCLUDED THE DAMAGE MAY BE DUE TO MISHANDLING / MISUSE. NO OTHER DAMAGE WAS FOUND. THE ENDOWRIST INSTRUMENTS INSTRUCTIONS FOR USE (IFU) SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS: HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; TUBE ABRASIONS WITH MATERIAL REMOVED ,FOUND DURING FAILURE ANALYSIS EVALUATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE, A LARGE SUTURECUT NEEDLE DRIVER INSTRUMENT CLICKS WHEN YOU TRY TO OPEN THE JAWS. THERE WERE NO MISSING PIECES OR FALLEN PIECES REPORTED. NO PATIENT HARM, ADVERSE OUTCOME, OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
577589 LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420296-02 M10121101 233

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM INSTRUMENTS AND ACCESSORIES