FDA Adverse Event Summary report: N

CENTURY 3

MDR report key: 34559 · Received May 24, 1996

Report

Report Number
34559
Date Received
May 24, 1996
Date of Event
May 4, 1996
Report Date
May 6, 1996
Manufacturer
COBE RENAL CARE, INC.
Product Code
FII
Report Source
User Facility report
Reporter Location
CO, US
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

WHILE TREATMENT WAS IN PROGRESS THE MACHINE ALARMED AIR IN BLOOD. UPON INSPECTION MICROBUBBLES WERE NOTED IN THE VENOUS SIDE OF THE PT'S CATHETER AND FOAM WAS NOTED IN THE VENOUS CHAMBER OF THE CARTRIDGE. PT C/O SOB, CHEST PAIN, DIAPHORESIS, LOW BP AND TACHYCARDIA. PT PLACED ON LEFT SIDE IN TRENDELBURG POSITION, TREATMENT DISCONTINUED, DR. NOTIFIED, O2 AND SALINE GIVEN. PT TRANSPORTED TO LOCAL HOSP WHERE HE WAS EVALUATED AND RELEASED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 CENTURY 3 C3 FII COBE RENAL CARE, INC. CS3 NA

Patients

Seq Age Sex Outcome Treatment
1 72 YR