FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3455588 · Received November 7, 2013

Report

Report Number
2531779-2013-19983
Event Type
Malfunction
Date Received
November 7, 2013
Report Date
October 25, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/25/2013 WITH THE FOLLOWING FINDINGS: THE PUMP WAS RETURNED WITH MOISTURE VISIBLE THRU THE DISPLAY LENS, THE OLED SCREEN IS BLANK DUE TO MOISTURE IN PUMP. THE PUMP BOOTED WITH AUDIO/VIBRATION. THERE WAS MOISTURE CORROSION VISIBLE IN THE BATTERY COMPARTMENT. THE BATTERY COMPARTMENT LEAK WAS FOUND DURING TESTING. THE PUMP COVER WAS REMOVED TO INSPECT INTERNAL COMPONENTS. MOISTURE CORROSION WAS VISIBLE IN PUMP COMPARTMENT. A VISUAL INSPECTION OF BATTERY COMPARTMENT REVEALED, COMPARTMENT CRACK FROM THREADS TO CASE SEAL. THERE WAS A SCRATCHED DISPLAY LENS FOUND AND THE CONTRAST AND OK SYMBOLS WERE WORN. (B)(4).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THE BATTERY COMPARTMENT IS CRACKED AND THE DISPLAY LENS WAS SCRATCHED. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/25/2013.