FDA Adverse Event
Other
Summary report: N
VANGUARD BREAST MRI AUXILIARY TABLE WITH 8CH
MDR report key: 3452895
·
Received October 25, 2013
Report
- Report Number
- 3005748597-2013-00003
- Event Type
- Other
- Date Received
- October 25, 2013
- Date of Event
- October 11, 2013
- Report Date
- October 24, 2013
- Manufacturer
- SENTINELLE MEDICAL, INC.
- Product Code
- MOS
- PMA / PMN Number
- K060873
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE HEALTHCARE FIELD ENGINEER (FE) WAS DISPATCHED TO AFFECTED SITE. THE FE REPORTED THAT THE LEFT DRAWER SLIDE HAD PULLED OVER ITS MECHANICAL STOP AND THE RIGHT DRAWER SLIDE WAS BENT AND HAD COMPLETELY BROKEN APART. BOTH SLIDE ASSEMBLIES TO BE REPLACED.
Description of Event or Problem · 1
THIS ADVERSE EVENT WAS REPORTED BY THE GE CLINICAL PRODUCT SURVEILLANCE SPECIALIST ((B)(4)) AS SUCH: "ON THURSDAY THE LEAD TECH PULLED THE SENTINELLE MAMMO TABLE'S LOWER ACCESSORY DRAWER OUT AND IT FELL ON HER FOOT. HER FOOT WAS INJURED BUT DID NOT REQUIRE MEDICAL TREATMENT." GE HEALTHCARE HAS CONTACTED THE SITE 3 ADD'L TIMES TO OBTAIN MORE INFO ON THE EXTENT OF THE FOOT INJURY AND TO SEE IF PT FOLLOW-UP IS REQUIRED. SITE HAS NOT RESPONDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 548403 | VANGUARD BREAST MRI AUXILIARY TABLE WITH 8CH | COIL, MAGNETIC RESONANCE, SPECIALTY | MOS | SENTINELLE MEDICAL, INC. | 4000001-11 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |