FDA Adverse Event Injury Summary report: N

INFUSION PUMP

MDR report key: 3449882 · Received November 5, 2013

Report

Report Number
3007566237-2013-03586
Event Type
Injury
Date Received
November 5, 2013
Date of Event
June 20, 2013
Report Date
October 7, 2013
Manufacturer
MEDTRONIC NEUROMODULATION
Product Code
LKK
PMA / PMN Number
P860004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

PRODUCT ID: NEU_UNKNOWN_CATH, LOT# UNKNOWN, PRODUCT TYPE: CATHETER. (B)(4).

Description of Event or Problem · 1

TEODORCZYK, J., SZMUDA, T., SIEMINSKI, M., LASS, P., STONIEWSKI, P. EVALUATION OF USEFULNESS OF SCINTIGRAPHIC IMAGING IN DIAGNOSIS OF INTRATHECAL DRUG DELIVERY SYSTEM MALFUNCTION ¿ A PRELIMINARY REPORT. POLISH JOURNAL OF RADIOLOGY. 2013. 78(3). DOI: 10.12659/PJR.889130. SUMMARY: SCINTIGRAPHIC STUDIES WERE PERFORMED IN SELECTED PATIENTS WITH NEUROLOGICAL DISEASES ASSOCIATED WITH SPASTICITY, WHO HAD IDDS SYSTEM IMPLANTED AND WERE UNABLE TO MAINTAIN SATISFYING CLINICAL EFFECT OF INRATHECALLY INFUSED BACLOFEN. AFTER EMPTYING THE IDDS SYSTEM OF THE DRUG, RADIOTRACER (99MTC-DTPA) SOLUTION WAS INJECTED INTO THE PUMP RESERVOIR. SUBSEQUENTLY, A SERIES OF SCINTIGRAPHIC IMAGES WAS REGISTERED, DEMONSTRATING PASSAGE AND DISTRIBUTION OF THE INFUSED RADIOTRACER. IN ALL INVESTIGATED CASES, SCINTIGRAPHIC STUDY RESULTED IN ACQUIRING RELEVANT ADDITIONAL DIAGNOSTIC INFORMATION. NORMAL OR DISRUPTED DISTRIBUTION OF RADIOTRACER IN SPINAL CANAL ALLOWED FOR A DIAGNOSIS DRUG RESISTANCE OR DEMONSTRATED PRESENCE OF ARACHNOID ADHESIONS RESPECTIVELY. EARLY APPEARANCE OF RADIOTRACER IN BLOOD WAS CONSIDERED A PROOF OF LEAK. OUR EXAMINATIONS HAD DECISIVE INFLUENCE ON FURTHER PATIENT TREATMENT, ALLOWING FOR DIAGNOSIS OF DRUG RESISTANCE IN ONE PATIENT OR COMPLICATION RELATED TO IDDS SYSTEM IN THREE OTHER CASES INCLUDING BREAKAGE OF A CATHETER, PUMP MALFUNCTION AND ARACHNOID ADHESIONS. SCINTIGRAPHIC METHODS CARRY SIGNIFICANT AMOUNT OF INFORMATION FACILITATING FINAL DIAGNOSIS OF THE CAUSE OF IDDS THERAPY FAILURE. THEY SHOULD BECOME AN IMPORTANT ELEMENT COMPLEMENTING THE DIAGNOSTIC STRATEGY IN PATIENTS WITH SUSPECTED FAILURE OF INTRATHECAL DRUG ADMINISTRATION SYSTEMS. INTERPRETATION OF RADIOISOTOPIC STUDIES, SINCE THEY ARE PURELY FUNCTIONAL, MUST BE PERFORMED IN STRICT RELATION TO CLINICAL DATA AND REPORTED EVENT: THIS REPORT REFERS TO A (B)(6) OLD FEMALE PATIENT. INTRATHECAL BACLOFEN (ITB) SYSTEM IMPLANTED TWO YEARS BEFORE. THE PATIENT RECEIVED INTRATHECAL BACLOFEN (MANUFACTURER UNKNOWN) FOR TREATMENT OF HUNTINGTON'S CHOREA. THE PATIENT HAD BEEN COMPLAINING OF WEAKENING OF THERAPEUTIC EFFECT FOR A YEAR. SCINTIGRAPHY REVEALS PROPER DISTRIBUTION OF RADIOTRACER IN CATHETERS AND GSE MALFUNCTION OF THE PUMPING ELEMENT WAS FINALLY DIAGNOSED, TAKING INTO CONSIDERATION THE INCREASED VOLUME OF RESIDUAL DRUG DURING PERIODIC REFILLS. DISRUPTION OF PUMP FUNCTION WAS CONFIRMED BY MANUFACTURER, RESULTING IN REPLACEMENT. AUTHOR CONCLUDED THAT RADIOLOGIC ASSESSMENT OF INTRATHECAL DRUG DELIVERY SYSTEMS (IDDS) SYSTEM WAS ESSENTIAL. IT FACILITATES FINAL DIAGNOSIS IN MANY CASES, WAS BROADLY AVAILABLE AND CARRIES BASIC MORPHOLOGIC INFORMATION NEEDED FOR INTERPRETATION OF FUNCTIONAL RADIOISOTOPIC STUDIES OR LATER SURGICAL INTERVENTION. SCINTIGRAPHIC IMAGING SEEMED TO BE A VALUABLE FUNCTIONAL ADJUNCT TO FIRST-LINE MORPHOLOGICAL STUDIES ENABLING FINAL DIAGNOSIS IN MANY PATIENTS WITH POTENTIAL MALFUNCTION OF INTRATHECAL DRUG DELIVERY SYSTEMS (IDDS). FURTHER STUDIES ENCOMPASSING HIGHER NUMBER OF PATIENTS WERE NECESSARY TO DEFINE THE EXACT DIAGNOSTIC ACCURACY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
568327 INFUSION PUMP PUMP, INFUSION, IMPLANTED, PROGRAMMABLE LKK MEDTRONIC NEUROMODULATION NEU_UNKNOWN_PUMP

Patients

Seq Age Sex Outcome Treatment
1 00029 YR Required Intervention