ABDR SSP UNITRAY KIT
Report
- Report Number
- 2244574-2013-00028
- Event Type
- Malfunction
- Date Received
- October 10, 2013
- Date of Event
- July 12, 2013
- Report Date
- July 12, 2013
- Manufacturer
- LIFE TECHNOLOGIES CORP.
- Product Code
- MZI
- PMA / PMN Number
- BK000019
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
A DISCREPANCY IN INTERNAL SOFTWARE SSP TEST DOCUMENTS RELATING TO PRIMER (B)(4) WAS IDENTIFIED. AS A RESULT, LABELING (AMBIGUITY LIST, UNITRAY AND ASG GEL DOCS, PM TABLES,. UCHFILES AND WORKSHEETS WILL BE CORRECTED. THE UPDATED REVISION WILL INCLUDE THE MIX SPECIFICITY TO REMOVE ALLELES DRB1 11:09, 11:83, 11:89, 11:113, 13:05:02, 13:06, 13:26:02, 13:42, 13:56, AND 13:136 AS POSITIVE. ADDITIONAL INVESTIGATION ACTIVITIES ARE STILL ONGOING AND WILL BE REPORTED IN THE 30-DAY REPORT.
UPON THE INTERNAL INVESTIGATION OF CUSTOMER COMPLAINT RECORD ((B)(4)) IT WAS DETERMINED THAT ONE LOT OF ABDR SSP UNITRAY KIT (CATALOG# 7810010) WAS AFFECTED. IN THE AFFECTED KIT LOTS, THE PRESENCE OF DRB1 13:05 IN A SAMPLE LEADS TO A MISTYPE OF DRB1 13:04 DUE TO A FALSE NEGATIVE RESULT IN LANE 15. THE REACTION PATTERNS FOR THESE TWO ALLELES IS AS FOLLOWS: DRB1 13:05: POSITIVE LANES 10, 11, 15, 26. DRB1 13:04: POSITIVE LANES 10, 11, 26.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 519976 | ABDR SSP UNITRAY KIT | MZI TEST, QUALITATIVE FOR HLA, NON-DIAGN | MZI | LIFE TECHNOLOGIES CORP. | 036 1316325 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |