INTERSTIM II

MDR report key: 3443310 · Received November 1, 2013

Report

Report Number
3004209178-2013-20049
Event Type
Malfunction
Date Received
November 1, 2013
Report Date
October 10, 2013
Manufacturer
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
Product Code
EZW
PMA / PMN Number
P970004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CONCOMITANT PRODUCTS: PRODUCT ID 3889-28, LOT # V796194, IMPLANTED: (B)(6) 2012, PRODUCT TYPE LEAD. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT NEVER HAD THERAPEUTIC EFFECT. NO STIMULATION SENSATION WAS NOTED. THERE WAS AN ISSUE REGARDING IMPEDANCE. THERE WAS READING >4000 OHMS ON SOME OF THE BIPOLAR PAIRS. THE PATIENT HAD BEEN HAVING INCREASED LEAKING AND NOT FEELING STIMULATION FOR THE PAST 3 WEEKS. NO REPORTING OF FALL, TRAUMA OR ANY MEDICAL PROCEDURE DONE. DEFAULT ELECTRODE IMPEDANCE TESTED AT 1V/210PW: C0: 1160 OHMS C1: 1087 OHMS C2 :1087 OHMS C3 : 535 OHMS 01: >4K OHMS 02: >4K OHMS 03: 1450 OHMS 12: 1122 OHMS 13: 1122 OHMS 23: 1087 OHMS RETEST ELECTRODE IMPEDANCE AT 1.5V/300PW: C0: 1160 OHMS C1: 1067 OHMS C2: 1087 OHMS C3: 535 OHMS 01: >4K OHMS 02: >4K OHMS 03: 1450 OHMS 12: 1122 OHMS 13: 1122 OHMS 23: 1087 OHMS THE PATIENT WAS PROGRAMMED ON PROGRAM 2 AT 5.8V AND WAS NOT FEELING ANY STIMULATION AND CONTINUED TO HAVE LEAKAGE FOR THE PAST 3 WEEKS. INS BATTERY LONGEVITY ESTIMATED AT 9-16 MONTHS REPROGRAMMING DONE 210PW/14HZ C+3-: 6.5V: NO STIM PROGRAM 1. 3-0+: 3.0V COMFORT IN PERINEUM AREA. PROGRAM 2. 2-3+ 3.0V COMFORT IN PERINEUM AREA. PROGRAM 3. 1-3+ 4.1V COMFORT IN PERINEUM AREA C+1- 4.5V NOTHING. 5.1V COMFORT IN PERINEUM AREA. PROGRAM 4. C+2- 3.5V COMFORT IN PERINEUM AREA. IF ADDITIONAL INFORMATION IS RECEIVED, A FOLLOW UP REPORT WILL BE SENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
564425 INTERSTIM II STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCONTINENCE EZW MEDTRONIC MED REL MEDTRONIC PUERTO RICO 3058

Patients

Seq Age Sex Outcome Treatment
1 00080 YR