FDA Adverse Event Malfunction Summary report: N

BOWEL GRASPER INSTRUMENT

MDR report key: 3441318 · Received October 31, 2013

Report

Report Number
2955842-2013-04843
Event Type
Malfunction
Date Received
October 31, 2013
Date of Event
October 3, 2013
Report Date
October 7, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE GRIP CABLES WERE BROKEN AT THE WRIST. THE INSTRUMENT WAS NON-INTUITIVE AS A RESULT. FAILURE ANALYSIS ALSO OBSERVED THAT THE MAIN TUBE INSULATION WAS SCRATCHED AND SOME OF THE INSULATION HAD BEEN REMOVED. THE SCRATCHES MEASURED APPROXIMATELY .023 - .114 IN LENGTH. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI SI TOTAL BENIGN HYSTERECTOMY PROCEDURE THE BOWEL GRASPER INSTRUMENT CABLE BROKE AT THE DISTAL END OF THE INSTRUMENT. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED AND NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
560853 BOWEL GRASPER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420177-04 S10121231 656

Patients

Seq Age Sex Outcome Treatment
1 43 YR DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES