ADULT UNHEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00854
- Event Type
- Malfunction
- Date Received
- October 31, 2013
- Date of Event
- September 27, 2013
- Report Date
- October 4, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BZO
- PMA / PMN Number
- K033710
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT RT134 ADULT UNHEATED BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR INSPECTION. IT WAS VISUALLY INSPECTED, PRESSURE TESTED, AND IMMERSED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: THE PRESSURE TEST RESULT WAS OUT OF SPECIFICATION DUE TO EXCESSIVE LEAK. WHEN THE SUBJECT BREATHING CIRCUIT WAS IMMERSED IN THE WATER BATH, THE LEAK WAS OBSERVED AT THE CONNECTION BETWEEN THE LID AND THE BOWL OF THE EXPIRATORY WATER TRAP. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 110905. CONCLUSION: THE BREATHING CIRCUIT WATER TRAP CONSISTS OF A BOWL AND A LID, WHICH CAN BE SEPARATED TO ALLOW THE CAREGIVER TO EMPTY THE WATER TRAP. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS ANY LEAK MUST HAVE DEVELOPED AFTER THE SUBJECT RT134 ADULT BREATHING CIRCUIT WAS RELEASED FOR DISTRIBUTION, DURING TRANSPORT, STORAGE OR USE, POSSIBLY BY DISTORTION OF THE WATER TRAP WHEN THE BOWL WAS CONNECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT134 ADULT UNHEATED BREATHING CIRCUIT STATE THE FOLLOWING: - "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." - "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." - "SET APPROPRIATE ALARMS."
A HOSPITAL IN (B)(6) REPORTED TO A DISTRIBUTOR THAT AN RT134 ADULT UNHEATED BREATHING CIRCUIT FAILED THE VENTILATOR LEAK TEST. THIS WAS OBSERVED BEFORE USE ON A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 559381 | ADULT UNHEATED BREATHING CIRCUIT | BZO | BZO | FISHER & PAYKEL HEALTHCARE LIMITED | RT134 | 110905 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |