FDA Adverse Event
Malfunction
Summary report: N
THORACIC GRASPER INSTRUMENT
MDR report key: 3438807
·
Received October 30, 2013
Report
- Report Number
- 2955842-2013-04740
- Event Type
- Malfunction
- Date Received
- October 30, 2013
- Date of Event
- October 3, 2013
- Report Date
- October 3, 2013
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE MAIN TUBE INSULATION WAS SCRATCHED AND SOME OF THE INSULATION HAD BEEN REMOVED. THE SCRATCHES MEASURED APPROXIMATELY .023 - .082 IN LENGTH. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE THE THORACIC GRASPER INSTRUMENT INSULATION WAS NOTED TO BROKEN AT THE DISTAL TIP OF THE INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 557897 | THORACIC GRASPER INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420343-01 | S10130204 790 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES |