FDA Adverse Event Malfunction Summary report: N

THORACIC GRASPER INSTRUMENT

MDR report key: 3438807 · Received October 30, 2013

Report

Report Number
2955842-2013-04740
Event Type
Malfunction
Date Received
October 30, 2013
Date of Event
October 3, 2013
Report Date
October 3, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS FOUND THAT THE MAIN TUBE INSULATION WAS SCRATCHED AND SOME OF THE INSULATION HAD BEEN REMOVED. THE SCRATCHES MEASURED APPROXIMATELY .023 - .082 IN LENGTH. NO OTHER DAMAGE FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER, THE REPORTED MALFUNCTION IF TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT PRIOR TO STARTING A DA VINCI SI SURGICAL PROCEDURE THE THORACIC GRASPER INSTRUMENT INSULATION WAS NOTED TO BROKEN AT THE DISTAL TIP OF THE INSTRUMENT. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
557897 THORACIC GRASPER INSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420343-01 S10130204 790

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI SI SYSTEM, INSTRUMENTS AND ACCESSORIES