AB/DR/DQ SSP UNITRAY
Report
- Report Number
- 2244574-2013-00024
- Event Type
- Malfunction
- Date Received
- October 2, 2013
- Date of Event
- August 23, 2013
- Report Date
- August 30, 2013
- Manufacturer
- LIFE TECHNOLOGIES CORP.
- Product Code
- MZI
- PMA / PMN Number
- BK000019
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE INTERNAL INVESTIGATION IDENTIFIED THAT THE REACTIVITY ASSIGNMENT IN OUR HUMAN OLIGOTYPING SOFTWARE (HOS) PROGRAM FOR A32:04 ALLELE IN PRIMER MIX PM019E WAS DETERMINED TO BE INCORRECT. PRODUCT LABELING WITH PRIMER MIX PM019E WILL BE UPDATED TO REMOVE SPECIFICITY FOR THE A 32:04 ALLELE. ADDITIONAL INVESTIGATION ACTIVITIES ARE STILL ONGOING AND WILL BE REPORTED IN THE 30-DAY REPORT. AFFECTED LOTS OF AB/DR/DQ SSP UNITRAY (CATALOG# 7850010). LOT#: 029 1414737, MFG. DATE: 2013-08, EXP. DATE: 2015-03.
(B)(4). THE REACTIVITY ASSIGNMENT IN OUR HUMAN OLIGOTYPING SOFTWARE (HOS) PROGRAM FOR A32:04 ALLELE IN PRIMER MIX PM019E WAS DETERMINED TO BE INCORRECT. TWO LOTS OF AB/DR/DQ SSP UNITRAY (CATALOG# 7850010) WERE AFFECTED BY THE INCORRECT A32:04 ALLELE ASSIGNMENT. THE KITS WILL PRODUCE AN INCORRECT LOW RESOLUTION TYPING OR MISTYPING RESULT OF A 02, A 03 ALLELES WHEN TESTING A SAMPLE WHICH HAS THE A 32:04 ALLELE. THE USER WILL BE UNAWARE OF THE INCORRECT TYPE, UNLESS THEY CONFIRM THE SAMPLE TYPING BY ANOTHER METHOD, THEREFORE, THIS WOULD BE CONSIDERED A MISTYPE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 498763 | AB/DR/DQ SSP UNITRAY | MZI TEST, QUALITATIVE FOR HLA, NON-DIAGN | MZI | LIFE TECHNOLOGIES CORP. | 029 1255127 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |