FDA Adverse Event Malfunction Summary report: N

ANALYTICAL E MODULE

MDR report key: 3429056 · Received October 25, 2013

Report

Report Number
1823260-2013-06570
Event Type
Malfunction
Date Received
October 25, 2013
Date of Event
October 8, 2013
Report Date
December 2, 2013
Manufacturer
ROCHE DIAGNOSTICS
Product Code
JJE
PMA / PMN Number
K961481
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

IT WAS UNKNOWN IF THE INITIAL REPORTER SENT REPORT TO THE FDA. (B)(4).

Additional Manufacturer Narrative · 1

A SPECIFIC ROOT CAUSE COULD NOT BE DETERMINED. NO HARDWARE FAILURE WAS DETERMINED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THEY RECEIVED ERRONEOUS RESULTS FOR ONE PATIENT SAMPLE TESTED FOR MULTIPLE ANALYTES ON 4 DIFFERENT ANALYZERS: 2 P MODULES, 1 E MODULE, AND 1 CORE MODULE. ON THIS E MODULE, THERE WAS AN ERRONEOUS RESULT FOR THYROTROPIN (TSH) THAT WAS REPORTED OUTSIDE OF THE LABORATORY. THE SAMPLE INITIALLY RESULTED AS 5.71 UIU/ML AND THIS VALUE WAS REPORTED OUTSIDE OF THE LABORATORY WHERE IT WAS QUESTIONED BY A PHYSICIAN. THE SAMPLE WAS REPEATED ON A DIFFERENT E MODULE ON (B)(6) 2013 WHERE IT RESULTED AS 2.64 UIU/ML. THE SAMPLE WAS ALSO REPEATED A SECOND TIME ON THE ORIGINAL E MODULE ON (B)(6) 2013, RESULTING AS 2.57 UIU/ML. THE REPEAT RESULT WAS BELIEVED TO BE CORRECT. THE PATIENT WAS NOT ADVERSELY AFFECTED. THE CUSTOMER WAS ASKED, BUT DID NOT PROVIDE THE TSH REAGENT LOT NUMBER OR EXPIRATION DATE. THE FIELD SERVICE REPRESENTATIVE COULD NOT DETERMINE A CAUSE. HE RAN DIAGNOSTICS. HE CHECKED PROBES AND PROBE BATHS; THESE CHECKED OK. HE CHECKED THE GEAR PUMP PRESSURE AND THIS WAS OK. HE RAN A PRECISION STUDY AND THIS WAS OK. THE CUSTOMER RAN QUALITY CONTROLS AND ALL RECOVERED WITHIN RANGE. THE SYSTEM WAS FOUND TO BE OPERATIONAL.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
548113 ANALYTICAL E MODULE IMMUNOCHEMISTRY ANALYZER JJE ROCHE DIAGNOSTICS NA NA

Patients

Seq Age Sex Outcome Treatment
1