FDA Adverse Event Malfunction Summary report: N

AIRFLOW

MDR report key: 3424418 · Received September 17, 2013

Report

Report Number
3424418
Event Type
Malfunction
Date Received
September 17, 2013
Date of Event
September 11, 2013
Report Date
September 17, 2013
Manufacturer
VENTLAB CORPORATION
Product Code
BTM
Product Problem
Yes
Report Source
User Facility report
Reporter Location
CT, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

WE BECAME AWARE OF AN ISSUE BY A SYSTEM HOSPITAL THAT THEY FOUND A "DUCK VALVE" THAT GOT STUCK WHILE TRYING TO USE A RESUSCITATION BAG VALVE MASK- WE CHECKED OUR SUPPLIES AND FOUND 28 DEFECTIVE VALVES. WITH SOME EXTRA PRESSURE THE VALVE WILL RELEASE, BUT IN A CODE SITUATION THESE VALVES COULD POSE A RISK. LOTS INVOLVED INCLUDED LOT # 106564 MANUFACTURED 3/13; LOT # 106003 MANUFACTURED 2/13; LOT # 105157 MANUFACTURED 1/13; LOT # 107509 MANUFACTURED 5/13; LOT # 99301 MANUFACTURED 1/13; LOT # 101304 MANUFACTURED 5/12; LOT # 98820 MANUFACTURED 12/11; LOT # 86341 MANUFACTURED 12/09; LOT # 85521 MANUFACTURED 11/09; LOT # 83709 MANUFACTURED 8/09; LOT # 79475 MANUFACTURED 12/08 AND LOT # 76380 MANUFACTURED 8/08.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
468840 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 106564
468841 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 106003
468842 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 105157
468843 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 107509
468844 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 99301
468845 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 101304
468846 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 98820
468847 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 86341
468848 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 85521
468849 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 83709
468850 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 76380
468982 AIRFLOW VENTILATOR, EMERGENCY, MANUAL BTM VENTLAB CORPORATION AF1000 79475

Patients

Seq Age Sex Outcome Treatment
1 *