FDA Adverse Event Malfunction Summary report: N

NOVA STATSTRIP

MDR report key: 3424231 · Received September 12, 2013

Report

Report Number
3424231
Event Type
Malfunction
Date Received
September 12, 2013
Date of Event
August 22, 2013
Report Date
September 12, 2013
Manufacturer
NOVA BIOMEDICAL CORP.
Product Code
FCO
Product Problem
Yes
Report Source
User Facility report
Reporter Location
TN, US
Reporter Occupation
INVALID DATA

Narratives

Description of Event or Problem · 1

AN ASSESSMENT WAS DONE OF METERS THROUGHOUT THE FACILITY. IT WAS NOTED THAT 14 OF THE 27 METERS HAD BATTERIES THAT WERE BULGING AND WOULD NOT ALLOW THE DOOR TO CLOSE PROPERLY ON THE METER. A NOVA CLINICAL SPECIALIST HELPED THE FACILITY IDENTIFY THIS PROBLEM. ALL OF THE BATTERIES ARE SUPPLIED THROUGH NOVA BIOMEDICAL, HOWEVER, THEY ARE MANUFACTURED BY FORMOSA ELECTRONIC INDUSTRIES, INC. (RECHARGEABLE LI-POLYMER BATTERY PACK, DC 3.7V). ALSO, THE FACILITY USES ANOTHER BATTERY MANUFACTURED BY A&S ENERGY CO., LTD. (MODEL AS2001, DC 3.7V, EXPIRATION DATE 11/13) HOWEVER THESE BATTERIES DO NOT BULGE AT ALL. WHEN THE METERS WERE FIRST IMPLEMENTED IN SPRING OF 2013, NOVA DID NOT INFORM THE FACILITY THAT THERE HAD BEEN PROBLEMS WITH THE BATTERIES AND THAT THE BATTERIES HAVE EXPIRATION DATES. ALSO OF NOTE IS THAT THE METERS WITH SWOLLEN BATTERIES SHOW THE LIGHTS ON THE RIGHT SIDE OF THE DOCKING STATION TO BE AMBER IN COLOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458968 NOVA STATSTRIP GLUCOSE METER, BATTERY FCO NOVA BIOMEDICAL CORP. AS2001 *

Patients

Seq Age Sex Outcome Treatment
1 * NOVA STASTSTRIP GLUCOSE METERS