FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3421428 · Received October 21, 2013

Report

Report Number
2531779-2013-18329
Event Type
Malfunction
Date Received
October 21, 2013
Report Date
October 9, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/09/2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE AUDIO BOLUS BUTTON WAS UNRESPONSIVE. THE COVER WAS REMOVED AND THE CONTACT WAS MISSING. ADDITIONALLY, THE CONTRAST KEY CONTACT WAS MISSING. ALL OF THE KEYPAD BUTTONS WERE UNRESPONSIVE DURING TESTING. THE KEYPAD COVER WAS LIFTING FROM THE PUMP. EVIDENCE OF CONTAMINATION WAS FOUND UNDER ALL KEY CONTACTS AND THE UP ARROW, DOWN ARROW, AND OK CONTACTS WERE MISALIGNED. THE CONTRAST KEY CONTACT WAS MISSING. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED A UNRESPONSIVE AUDIO BOLUS BUTTON. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/09/2013. THERE WAS NO ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
538900 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 17 YR