FDA Adverse Event Other Summary report: N

PRISMALIX W/ ENERGIX

MDR report key: 3420976 · Received September 24, 2013

Report

Report Number
9710055-2013-00034
Event Type
Other
Date Received
September 24, 2013
Manufacturer
MAQUET SAS
Product Code
FSY
PMA / PMN Number
K040015
Removal / Correction Number
NA
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A MAQUET FIELD SERVICE TECHNICIAN (FST) INSPECTED THE DEVICE AND FOUND THAT THE RETAINING CLIP (A.K.A.CIRCLIP) USED TO SECURE THE LIGHTING SYSTEM TO ITS DOWN TUBE MOUNT, HAD BEEN INCORRECTLY INSTALLED. THE CIRCLIP WAS BENT AND OUT OF ITS GROOVE. THE MAQUET INVESTIGATION REVEALED THAT THE ARM HAD BEEN DISMANTLED IN (B)(4) 2013 DURING MAINTENANCE DONE BY A THIRD PARTY MAINTENANCE PROVIDER. MAQUET DETERMINED THAT THE TECHNICIAN INCORRECTLY POSITIONED THE CIRCLIP AT THE END OF THIS MAINTENANCE WHICH LED TO THE REPORTED EVENT. THE MAQUET FST REPAIRED THE DEVICE, INSTALLING A NEW CIRCLIP AND WASHERS, AND RETURNED THE UNIT TO SERVICE. THE PRISMALIX SERIES INSTALLATION MANUAL INCLUDES ALL NECESSARY INSTRUCTIONS TO INSTALL THE ARM TO THE CEILING TUBE. (B)(4). MAQUET MEDICAL SYSTEM USA SUBMITS THIS REPORT ON BEHALF OF THE DEVICE MANUFACTURING FACILITY. MAQUET (B)(4) PROVIDES PRODUCT FAILURE INVESTIGATION, ANALYSIS AND RESOLUTION FOR THE DEVICE DESCRIBED IN THIS REPORT.

Description of Event or Problem · 1

THE CUSTOMER REPORTED TO MAQUET THAT DURING A SURGICAL INTERVENTION, WHILE A NURSE WAS ATTEMPTING TO REPOSITION A CUPOLA, THE SURGICAL LIGHT DETACHED FROM THE CEILING MOUNTING TUBE. THE FALLING COMPONENTS STRUCK THE PATIENT ON HIS RIGHT SIDE FROM THE FOREARM TO THE CHIN. THE PATIENT WAS SENT FOR AN EXAMINATION. THE HOSPITAL HAS NOT PROVIDED ANY DETAILS WITH RESPECT TO THE PATIENT RESULTANT FROM THIS EVENT. THE SURGEON WAS STRUCK ON HIS HEAD, CONCUSSED AND REQUIRED TWO STITCHES. (B)(4). REF. IMP REPORT #3008355164-2013-00232.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
483068 PRISMALIX W/ ENERGIX FSY MAQUET SAS

Patients

Seq Age Sex Outcome Treatment
1