FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3420944 · Received October 21, 2013

Report

Report Number
2531779-2013-18275
Event Type
Malfunction
Date Received
October 21, 2013
Report Date
October 9, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/09/2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED KEYPAD PEELING FROM BOTTOM THROUGH "CONTRAST" KEY. A DAMAGED KEYPAD WILL ALLOW CONTAMINATION TO PERMEATE THE BUTTON CONTACTS NEGATIVELY IMPACTING THE BUTTON FUNCTION. THE DAMAGED KEYPAD IS OBVIOUS AND DETECTABLE AND SHOULD ALERT THE USER TO DISCONTINUE USE OF THE PUMP. THE CONTRAST KEY IS NOT RESPONDING. THE UP, DOWN AND OK KEYS RESPONDED APPROPRIATELY. THE KEYPAD FLEX CABLE WAS DAMAGED AT CONTRAST KEY. THERE WAS GREEN MOISTURE CONTAMINATION UNDER ALL KEY CONTACTS. THERE WERE BATTERY COMPARTMENT CRACKS OBSERVED IN THREAD AREA AND BELOW GRIP PAD. THERE WAS EVIDENCE OF MOISTURE CONTAMINATION INSIDE BATTERY COMPARTMENT. LEAK TEST SHOWED LEAK AT CRACKS IN BATTERY COMPARTMENT. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED THAT THERE WAS CONTAMINATION UNDER THE BUTTONS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/09/2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
539527 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 27 YR