FDA Adverse Event Death Summary report: N

ENDURANT

MDR report key: 3419311 · Received October 18, 2013

Report

Report Number
2953200-2013-02021
Event Type
Death
Date Received
October 18, 2013
Date of Event
December 6, 2013
Report Date
December 9, 2013
Manufacturer
MEDTRONIC IRELAND
Product Code
MIH
PMA / PMN Number
P100021
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
IS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION, RESULTS: INHERENT RISK OF PROCEDURE (OCCLUSION). (UNKNOWN CAUSE OF EVENT). CONCLUSION: (UNKNOWN CAUSE OF EVENT). (B)(4).

Additional Manufacturer Narrative · 1

(B)(4). RESULTS, CONCLUSION: ANATOMY RELATED; HYPERCOAGULABILITY.

Additional Manufacturer Narrative · 1

METHOD: FILM.

Additional Manufacturer Narrative · 1

METHOD: FILM. RESULTS, CONCLUSION: UNKNOWN CAUSE OF DEATH. PATIENT CONDITION; THROMBUS MAY HAVE BEEN DUE TO HYPERCOAGULABILITY.

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THE PATIENT EXPIRED. THE CAUSE OF DEATH IS UNKNOWN. THE INVESTIGATOR ASSESSED THE DEATH AS DEVICE AND PROCEDURE RELATED. FILM REVIEW ANALYSIS: CTA 3-YEARS POST-IMPLANT REVEALED THAT THE BIFUR WAS PLACED JUST BELOW THE RENALS, INTO A VERY SHORT LENGTH NECK. THE IPSILATERAL LIMB AND EXTENSION WAS POSITIONED WITHIN THE LEFT COMMON ILIAC, AND THE CONTRALATERAL AND EXTENSION WAS IN THE RIGHT COMMON ILIAC. THE AAA WAS BI-LOBAL; THE PROXIMAL MAX ANEURYSM DIAMETER WAS 5.5CM AND THE DISTAL MAX AAA DIAMETER WAS 10CM. AREAS OF THROMBUS WAS SEEN WITHIN THE DESCENDING THORACIC AORTA, AND WITHIN THE THORACO-ABDOMINAL AORTA. THE ENTIRE STENT GRAFT SYSTEM WAS SEEN PARTIALLY OCCLUDED BY THROMBUS. BEGINNING AT THE PROXIMAL STENT GRAFT MARGIN, THE BIFUR AORTIC BODY WAS APPROXIMATELY 50 PERCENT OCCLUDED DOWN TO THE FLOW DIVIDER, AND BOTH LIMBS/EXTENSIONS SHOWED AREAS OF THROMBUS ALONG THEIR ENTIRE LENGTH; PRIMARILY SEEN ALONG THE INNER WALL OF THE STENT GRAFTS. NEAR TO THE ORIGIN OF THE LIMBS, BOTH THE IPSI AND CONTRA LIMBS WERE ANGULATED POST APPROXIMATELY 70 DEGREE; BUT NO OBVIOUS KINKS OR LIMB COMPRESSION WAS SEEN. PRE-IMPLANT AND EARLIER POST-IMPLANT FILMS WERE NOT PROVIDED. THE CAUSE OF THE THORACIC AORTA AND WIDESPREAD STENT GRAFT THROMBUS COULD NOT BE DETERMINED AND MAY BE RELATED TO A PATIENT CONDITION (HYPERCOAGULABILITY). CTA AT 3 YEARS POST-IMPLANT, 1 WEEK AFTER THE REPORTED RESOLUTION OF THE THROMBUS USING ANTICOAGULANTS, REVEALED SIMILAR FINDINGS FROM THE STUDY 1-MONTH PRIOR. THE BI-LOBAL AAA DIAMETER MEASURED 5.9 CM MAX (PROXIMAL ANEURYSM) AND 11 CM MAX (DISTAL AAA); SLIGHTLY INCREASED FROM PREVIOUS STUDY. AGAIN, THROMBUS WAS SEEN WITHIN THE BIFURCATE AORTIC BODY AND ALONG THE LENGTH OF BOTH LIMBS. NO OBVIOUS STENT GRAFT KINKS OR COMPRESSION WAS SEEN. THE CAUSE OF THE THROMBUS COULD NOT BE DETERMINED AND MAY BE PATIENT RELATED, AND NO STENT GRAFT RELATED ISSUES TO THE PATIENT'S DEATH COULD BE DETERMINED.

Description of Event or Problem · 1

AN ENDURANT STENT GRAFT SYSTEM WAS IMPLANTED IN A PATIENT FOR THE ENDOVASCULAR TREATMENT OF A 110 MM IN DIAMETER FUSIFORM ABDOMINAL AORTIC ANEURYSM. LENGTH OF NON-ANEURYSM AORTIC NECK WAS 15 MM, PROXIMAL DIAMETER OF NON-ANEURYSM AORTIC NECK WAS 24 MM, DISTAL DIAMETER OF NON-ANEURYSMAL AORTIC NECK WAS 25 MM, DIAMETER OF RIGHT ILIAC ARTERY WAS 13 MM, DIAMETER OF LEFT ILIAC ARTERY WAS 9 MM, DIAMETER OF RIGHT FEMORAL ARTERY WAS 12 MM, AND DIAMETER OF LEFT FEMORAL ARTERY WAS 10 MM. THE RIGHT ILIAC ARTERY WAS MILDLY TORTUOUS AND THE LEFT ILIAC ARTERY WAS MODERATELY TORTUOUS. DEGREE OF CIRCUMFERENTIAL THROMBUS AND / OR CALCIFICATION WAS 30%. IT WAS REPORTED THAT APPROXIMATELY THREE YEARS POST INDEX PROCEDURE, THE PATIENT WAS HOSPITALIZED DUE TO ABDOMINAL PAIN. A CTA SCAN SHOWED THROMBUS IN THE ABDOMINAL AORTA. THE PATIENT WAS TREATED WITH ANTICOAGULANTS RESOLVING THE EVENT. THE PHYSICIAN ASSESSED THIS EVENT AS NOT DEVICE OR PROCEDURE RELATED. NO ADDITIONAL CLINICAL SEQUELAE WERE REPORTED.

Description of Event or Problem · 1

FILM REVIEW ANALYSIS: CTA IMAGES 3-YEARS POST-IMPLANT REVEALED THAT THE BIFURCATE WAS PLACED JUST BELOW THE RENALS, INTO A VERY SHORT LENGTH NECK. THE IPSILATERAL LIMB AND EXTENSION WAS POSITIONED WITHIN THE LEFT COMMON ILIAC, AND THE CONTRA LIMB AND EXTENSION WAS IN THE RIGHT COMMON ILIAC. THE AAA WAS BI-LOBAL; THE PROXIMAL MAX ANEURYSM DIAMETER WAS 5.5CM AND THE DISTAL MAX AAA DIAMETER WAS 10CM. AREAS OF THROMBUS WAS SEEN WITHIN THE DESCENDING THORACIC AORTA, AND WITHIN THE THORACO-ABDOMINAL AORTA. THE ENTIRE STENT GRAFT SYSTEM WAS SEEN PARTIALLY OCCLUDED BY THROMBUS. BEGINNING AT THE PROXIMAL STENT GRAFT MARGIN, THE BIFURCATE AORTIC BODY WAS APPROXIMATELY 50% OCCLUDED DOWN TO THE FLOW DIVIDER, AND BOTH LIMBS/EXTENSIONS SHOWED AREAS OF THROMBUS ALONG THEIR ENTIRE LENGTH; PRIMARILY SEEN ALONG THE INNER WALL OF THE STENT GRAFTS. NEAR TO THE ORIGIN OF THE LIMBS, BOTH THE IPSI AND CONTRA LIMBS WERE ANGULATED POSTERIORLY APPROXIMATELY 70DEG; BUT NO OBVIOUS KINKS OR LIMB COMPRESSION WAS SEEN. PRE-IMPLANT AND EARLIER POST-IMPLANT FILMS WERE NOT PROVIDED. THE CAUSE OF THE THORACIC AORTA AND WIDESPREAD STENT GRAFT THROMBUS COULD NOT BE DETERMINED AND MAY BE RELATED TO A PATIENT CONDITION (HYPERCOAGULABILITY).

Description of Event or Problem · 1

ADDITIONAL INFORMATION WAS RECEIVED. IT WAS REPORTED THAT THE LOCATION OF THE THROMBUS WAS LOCATED IN THE ABDOMINAL AORTA AND INSIDE THE STENT GRAFT. THE OCCLUSION/THROMBUS LOCATED IN CORONARY AND IN THE AORTA AS INDICATED IN THE CT CANNOT BE DEVICE RELATED AND PROBABLY DERIVES FROM HYPERCOAGULABILITY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
535388 ENDURANT SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH MEDTRONIC IRELAND V00508961

Patients

Seq Age Sex Outcome Treatment
1 00077 YR Death| R