FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3416157 · Received October 17, 2013

Report

Report Number
2531779-2013-17920
Event Type
Malfunction
Date Received
October 17, 2013
Report Date
October 2, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/02/2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THE KEYPAD WAS FULLY INTACT BUT THE KEYPAD WAS WORN. ALL OF THE BUTTONS RESPONDED APPROPRIATELY. EVALUATION REVEALED THERE WAS CONTAMINATION UNDER THE UP, DOWN AND CONTRACT KEY CONTACTS. UNRELATED TO THE CASE, THE DISPLAY LENS WAS SCRATCHED. (B)(6).

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED CONTAMINATION UNDER BUTTONS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/02/2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
534176 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR