- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 10/02/2013 WITH THE FOLLOWING FINDINGS: EVALUATION REVEALED THE KEYPAD WAS FULLY INTACT BUT THE KEYPAD WAS WORN. ALL OF THE BUTTONS RESPONDED APPROPRIATELY. EVALUATION REVEALED THERE WAS CONTAMINATION UNDER THE UP, DOWN AND CONTRACT KEY CONTACTS. UNRELATED TO THE CASE, THE DISPLAY LENS WAS SCRATCHED. (B)(6).
Description of Event or Problem · 1
THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED CONTAMINATION UNDER BUTTONS. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 10/02/2013.