FDA Adverse Event Injury Summary report: N

TVC INSIGHT CATHETER

MDR report key: 3412997 · Received October 10, 2013

Report

Report Number
3004722468-2013-00002
Event Type
Injury
Date Received
October 10, 2013
Date of Event
March 21, 2013
Report Date
October 10, 2013
Manufacturer
INFRAREDX INC.
Product Code
OGZ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

PATIENT EXPERIENCED VENTRICULAR FIBRILLATION DURING CATHETER DWELL TIME. VIA RIGHT FEMORAL ARTERY WITH SELDINGER TECHNIQUE AND ONE WALL STICK, THE RIGHT FEMORAL ARTERY WAS ENTERED AND A 6 FRENCH SHEATH WAS PLACED. A NUMBER 4 LEFT SELECTIVE CORONARY CATHETER WAS ADVANCED TO THE ASCENDING AORTA. THE LEFT CORONARY CATHETER SELECTIVELY CANNULATED. NUMEROUS INJECTIONS PERFORMED WITH DIGITAL FILMING. THIS CATHETER REPLACED BY NUMBER 4 RIGHT SELECTIVE CORONARY CATHETER. IT WAS ADVANCED TO THE ASCENDING AORTA, THE RIGHT CORONARY ARTERY SELECTIVELY CANNULATED. NUMEROUS INJECTIONS PERFORMED WITH JUST DIGITAL FILMING. REPLACED WITH PIGTAIL CATHETER. PRESSURES WERE RECORDED AND ALL CATHETERS WERE REMOVED. MOVED ON TO IMAGING AS WELL AS FRACTIONAL FLOW WIRE. DURING IMAGING CATHETER MAY HAVE OCCLUDED FLOW DUE TO STENOSIS. VENTRICULAR TACHYCARDIA FIBRILLATION DEVELOPED REQUIRING CARDIOVERSION PROMPTLY BACK TO BASELINE. PHYSICIAN REPORTED EVENT WAS DUE TO ISCHEMIA FROM LONG DWELL TIME - WOULD HAVE OCCURRED WITH ANY IVUS CATHETER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
518193 TVC INSIGHT CATHETER NIR/IVUS CORONARY IMAGING CATHETER OGZ INFRAREDX INC. NIRS-MC7-7-3.2F NA

Patients

Seq Age Sex Outcome Treatment
1 54 YR Life Threatening| R