FDA Adverse Event Other Summary report: N

WASP

MDR report key: 3410538 · Received September 20, 2013

Report

Report Number
3002444944-2013-00003
Event Type
Other
Date Received
September 20, 2013
Date of Event
August 22, 2013
Report Date
September 18, 2013
Manufacturer
COPAN ITALIA SPA
Product Code
JTC
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COMPLAINT WAS PROCESSED AS (B)(4) PROCEDURE AND THE INVESTIGATION AND MDR EVALUATION HAS BEEN STARTED. THE INCIDENT WITH THE MACHINE WAS CAUSED BY ACCIDENTALLY TOUCHING OF A HOT SURFACE INSIDE THE MACHINE BY THE OPERATOR. THE HOT SURFACE WAS INDIVIDUATED AS THE INCINERATOR EXTERNAL SIDE. THE INCINERATOR IS PRESUMED TO BE HOT BECAUSE ITS FUNCTION AS STERILIZER. THE INCINERATOR IS LOCATED INTERNALLY IN THE WASP MACHINE AND IT IS IDENTIFIED AS A HOT SURFACE BY THE WARNING SYMBOL LABELING. THE INCINERATOR IS REACHABLE FROM THE USER ONLY WHEN THE MACHINE IS STOPPED AND IN EMERGENCY STATE AND ONLY ACCIDENTAL CONTACT WITH THE HOT SURFACE COULD BE POSSIBLE. THE OPERATOR WAS INSTRUCTED TO NOT PASS OVER THE INCINERATOR AREA. TESTING AT THE LABORATORY SIDE BY (B)(4) TECHNICAL SUPPORT VERIFIED THAT NOT ABNORMAL SITUATION WAS IN THE MACHINE. A CAPA WAS OPENED IN ORDER TO REVISE THE USER MANUAL ADDING MORE WARNINGS ABOUT THE ACCESS TO THE AREA CLOSEST TO THE INCINERATOR.

Description of Event or Problem · 1

THE WASP IS AN AUTOMATIC EQUIPMENT FOR THE PLANTING AND STREAKING OF CLINICAL SPECIMENS. DURING THE ORDINARY MAINTENANCE OF THE WASP MACHINE S/N: (B)(4) INSTALLED IN THE (B)(6) HOSPITAL, THE OPERATOR ACCIDENTALLY TOUCHED THE HOT SURFACE OF THE EXTERNAL SIDE OF THE INCINERATOR CAUSING A RUBESCENCE IN HIS ARM. THE INCINERATOR IS USED FOR THE LOOP STERILIZATION AND IT IS LOCATED INTERNALLY TO THE WASP VISIBLY IDENTIFIED BY THE HOT SURFACE WARNING SYMBOL LABELING. NO MEDICAL INTERVENTION WAS REQUIRED AND NO EVIDENCE OF THE INJURY REMAINED. THE INCIDENT HAPPENED DURING THE OPERATOR TRAINING AND DESPITE OF TRAINER RECOMMENDATION TO NOT PASS OVER THE INCINERATOR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
476529 WASP JTC COPAN ITALIA SPA W086

Patients

Seq Age Sex Outcome Treatment
1