FDA Adverse Event Malfunction Summary report: N

COULTER® LH 780 HEMATOLOGY ANALYZER

MDR report key: 3399782 · Received October 10, 2013

Report

Report Number
1061932-2013-02212
Event Type
Malfunction
Date Received
October 10, 2013
Date of Event
September 14, 2013
Report Date
September 14, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061574
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A BEC FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2013 TO EVALUATE THE INSTRUMENT. THE FSE FOUND A HOLE IN THE TUBING THROUGH PINCH VALVE VL46A. THE FSE REPLACED THE TUBING AND THE INSTRUMENT RAN WITHOUT ANY LEAKS. THE REPAIRS WERE VERIFIED PER ESTABLISHED PROCEDURES. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER CONTACTED BECKMAN COULTER (BEC) TO REPORT A LEAK INSIDE THE COULTER LH 780 HEMATOLOGY ANALYZER. THE VOLUME OF THE LEAK WAS ABOUT 50 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF A LABORATORY COAT, GLOVES, AND PROTECTIVE EYEWEAR AT THE TIME OF OCCURRENCE. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THIS EVENT.