FDA Adverse Event Malfunction Summary report: N

LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT

MDR report key: 3392506 · Received October 7, 2013

Report

Report Number
2955842-2013-04100
Event Type
Malfunction
Date Received
October 7, 2013
Date of Event
August 22, 2013
Report Date
September 9, 2013
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K050369
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION WAS UNABLE TO CONFIRM THE CUSTOMER REPORTED FAILURE MODE. THE INSTRUMENT INSTALLED ON AN IN-HOUSE IS2000 SYSTEM PASSED A SUTURE CUT TEST. THE BLADES DID EXHIBIT ANY DAMAGE. FAILURE ANALYSIS INVESTIGATION ALSO FOUND THAT THE INSTRUMENT'S MAIN TUBE HAD VARIOUS SCRATCH MARKS WITH LIGHT MATERIAL REMOVAL. THE SCRATCHES WERE .061 - .226 IN LENGTH AND WERE NOT ALIGNED WITH THE TUBE AXIS. IT WAS CONCLUDED THAT THE DAMAGE TO THE INSTRUMENT'S MAIN TUBE WAS LIKELY DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE WAS FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS, HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. O DO NOT USE AN INSTRUMENT TO CLEAN DEBRIS FROM ANOTHER INSTRUMENT INTRAOPERATIVELY. THIS MAY RESULT IN DAMAGE TO THE INSTRUMENTS OR OTHER UNINTENDED CONSEQUENCES, SUCH AS DISCONNECTION OF THE INSTRUMENT TIP. THE CUSTOMER REPORTED COMPLAINT DOES NOT IN ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE DAMAGE TO THE INSTRUMENT'S MAIN TUBE, FOUND DURING FAILURE ANALYSIS INVESTIGATION COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A DA VINCI S SURGICAL PROCEDURE, THE SCISSORS ON THE LARGE SUTURECUT NEEDLE DRIVER INSTRUMENT WERE DULL. NO MISSING OR FALLEN PIECES WERE REPORTED. THE PLANNED SURGICAL PROCEDURE WAS COMPLETED. NO PATIENT HARM, ADVERSE OUTCOME OR INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
507293 LARGE SUTURECUT NEEDLE DRIVERINSTRUMENT ENDOSCOPIC INSTRUMENT NAY INTUITIVE SURGICAL,INC. 420296-02 M10120509 807

Patients

Seq Age Sex Outcome Treatment
1 DA VINCI S SURG SYS, INSTRUMENTS AND ACCESSORIES