ADULT UNHEATED BREATHING CIRCUIT

MDR report key: 3382122 · Received October 2, 2013

Report

Report Number
9611451-2013-00755
Event Type
Malfunction
Date Received
October 2, 2013
Date of Event
September 4, 2013
Report Date
September 6, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BZO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: ONE OF THE THREE RT134 ADULT UNHEATED BREATHING CIRCUITS WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR INSPECTION. IT WAS VISUALLY INSPECTED, PRESSURE TESTED, AND IMMERSED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: THE PRESSURE TEST RESULT WAS OUT OF SPECIFICATION DUE TO EXCESSIVE LEAK. WHEN THE SUBJECT BREATHING CIRCUIT WAS IMMERSED IN THE WATER BATH, THE LEAK WAS OBSERVED AT THE CONNECTION BETWEEN THE LID AND THE BOWL OF THE EXPIRATORY WATER TRAP. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 130426. CONCLUSION: THE BREATHING CIRCUIT WATER TRAP CONSISTS OF A BOWL AND A LID, WHICH CAN BE SEPARATED TO ALLOW THE CAREGIVER TO EMPTY THE WATER TRAP. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS ANY LEAK MUST HAVE DEVELOPED AFTER THE SUBJECT RT134 ADULT BREATHING CIRCUIT WAS RELEASED FOR DISTRIBUTION, DURING TRANSPORT, STORAGE OR USE, POSSIBLY BY DISTORTION OF THE WATER TRAP WHEN THE BOWL WAS CONNECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT134 ADULT UNHEATED BREATHING CIRCUIT STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE ALARMS." THE HOSPITAL PRESSURE TESTED THE SUBJECT BREATHING CIRCUIT AND DETECTED THE MALFUNCTION BEFORE CONNECTING TO THE PATIENT.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED TO A DISTRIBUTOR THAT THREE RT134 ADULT UNHEATED BREATHING CIRCUITS FAILED THE VENTILATOR LEAK TEST. THIS WAS OBSERVED BEFORE USE ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
497907 ADULT UNHEATED BREATHING CIRCUIT BZO BZO FISHER & PAYKEL HEALTHCARE LIMITED RT134 130426

Patients

Seq Age Sex Outcome Treatment
1