ADULT INSPIRATORY HEATED BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00751
- Event Type
- Malfunction
- Date Received
- October 1, 2013
- Date of Event
- September 5, 2013
- Report Date
- September 5, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- BTT
- PMA / PMN Number
- SEE H10
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RT105 IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: THE COMPLAINT BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE (B)(4) FOR EVALUATION. THE COMPLAINT CIRCUIT WAS PRESSURE TESTED TO OUR PRODUCT SPECIFICATION AND SUBMERGED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: THE PRESSURE TEST IDENTIFIED A LEAK AND THE WATER BATH TEST IDENTIFIED THE LEAK TO BE AT THE CONNECTION BETWEEN THE LID AND BOWL OF THE WATER TRAP, CONFIRMING THE REPORTED FAULT. A LOT CHECK REVEALED ONE OTHER COMPLAINT FOR LOT NUMBER 130515. CONCLUSION: THE RT105 BREATHING CIRCUIT WATER TRAP CONSISTS OF A BOWL AND A LID. THE WATER TRAP BOWL AND LID CAN BE SEPARATED TO ALLOW THE CAREGIVER TO EMPTY THE WATER TRAP. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS THAT ANY LEAK MUST HAVE DEVELOPED AFTER THE BREATHING CIRCUITS WERE RELEASED FOR DISTRIBUTION, DURING TRANSPORT, STORAGE OR USE, POSSIBLY BY DISTORTION OF THE WATER TRAP WHEN THE BOWL WAS CONNECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT105 ADULT BREATHING CIRCUIT STATE THE FOLLOWING: - "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." - "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." - "SET APPROPRIATE VENTILATOR ALARMS." THE HOSPITAL CORRECTLY FOLLOWED OUR USER INSTRUCTIONS AND DETECTED THE LEAKS DURING A SETUP CHECK BEFORE USE ON A PATIENT.
A HOSPITAL IN (B)(6) REPORTED TO A FISHER & PAYKEL HEALTHCARE FIELD REPRESENTATIVE THAT AN RT105 ADULT BREATHING CIRCUIT FAILED THE LEAK TEST ON THE SERVO-I VENTILATOR. THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 494559 | ADULT INSPIRATORY HEATED BREATHING CIRCUIT | BTT | BTT | FISHER & PAYKEL HEALTHCARE LIMITED | RT105 | 130515 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | SERVO-I VENTILATOR |