FDA Adverse Event Injury Summary report: N

TI MATRIX LOCKING CAP

MDR report key: 3375991 · Received September 30, 2013

Report

Report Number
2530088-2013-01190
Event Type
Injury
Date Received
September 30, 2013
Date of Event
December 14, 2012
Report Date
May 2, 2013
Manufacturer
SYNTHES BRANDYWINE
Product Code
NKB
PMA / PMN Number
K100952
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. ADDITIONAL PRODUCT CODE FOR THIS REPORT INCLUDES: MNH, MNI, KWO, KWP. INVESTIGATION COULD NOT BE COMPLETED AND NO CONCLUSION COULD BE DRAWN AS NO DEVICE WAS RETURNED AND NO LOT NUMBER OR PART NUMBER WAS PROVIDED. FURTHER REVIEW OF THE COMPLAINT WAS REQUIRED AND THE FOLLOWING PARTS WERE ADDED TO THE COMPLAINT THAT WERE NOT ORIGINALLY CAPTURED ON INITIAL FILING DATE (B)(4) 2013: ON (B)(6) 2013 - 3 OF 3 - TI MATRIX LOCKING CAP; PART NUMBER 04.632.000 WAS ADDED; LOT NUMBER WAS NOT PROVIDED.

Description of Event or Problem · 1

A PATIENT WAS IMPLANTED WITH THE MATRIX MIS SYSTEM FOR POSTERIOR LUMBAR INTERBODY FUSION (PLIF) ON (B)(6) 2012. ALL POLYAXIAL SCREWHEADS WERE REMOVED AND REDUCTION HEADS WERE PLACED ON THE SCREWS PRIOR TO IMPLANTATION. ON AN UNKNOWN DATE, PATIENT PRESENTS TO SURGEON FOR POST-OPERATIVE FOLLOWUP COMPLAINING OF PAIN. EXAMINATION AND X-RAY REVEAL THE RIGHT SIDE OF THE CONSTRUCT SHOWS THE L4 LOCKING CAP TO BE FREE FLOATING OUT OF THE POLYAXIAL HEAD, AND THE L5 AND S1 CAPS TO BE LOOSE BUT STILL ENGAGED IN THE POLYAXIAL HEADS OF THE SCREWS. THE ROD ITSELF HAS SLID DOWN AND PASSED THROUGH THE HEADS OF THE L4 AND L5 SCREWS AND IS ALMOST FREE FLOATING OVER THE SACRUM. ON AN UNKNOWN DATE PATIENT RETURNED TO OR FOR SURGICAL INTERVENTION. SURGEON REPOSITIONED ROD AND REPOSITIONED L4 LOCKING CAP, AND RETIGHTENED L4, L5, AND S1 LOCKING CAPS. THIS IS REPORT 5 OF 5 FOR COMPLAINT # (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
491263 TI MATRIX LOCKING CAP NKB SYNTHES BRANDYWINE

Patients

Seq Age Sex Outcome Treatment
1 50 YR Required Intervention