COULTER® LH 750 HEMATOLOGY ANALYZER

MDR report key: 3375129 · Received September 30, 2013

Report

Report Number
1061932-2013-02096
Event Type
Malfunction
Date Received
September 30, 2013
Date of Event
September 1, 2013
Report Date
September 1, 2013
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K061574
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
LA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER FOUND THE TUBING WHICH HAD POPPED OFF OF THE FITTING ON THE BLOOD SAMPLING VALVE (BSV). THE CUSTOMER RECONNECTED THE TUBING WHICH RESOLVED THE LEAK. SERVICE WAS NOT DISPATCHED FOR THIS EVENT. PER LABELING, BECKMAN COULTER URGES ITS CUSTOMERS TO COMPLY WITH ALL NATIONAL HEALTH AND SAFETY STANDARDS SUCH AS THE USE OF BARRIER PROTECTION. THIS MAY INCLUDE, BUT IT IS NOT LIMITED TO, PROTECTIVE EYEWEAR, GLOVES, AND SUITABLE LABORATORY ATTIRE WHEN OPERATING OR MAINTAINING THIS OR ANY OTHER AUTOMATED LABORATORY ANALYZER. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER (BEC) TO REPORT THAT THE COULTER LH 750 HEMATOLOGY ANALYZER WAS LEAKING 1 TO 2 ML OF DILUENT UNDER THE INSTRUMENT. THE LEAK WAS NOT CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT (PPE) CONSISTING OF LABORATORY COAT, GLOVES, AND PROTECTIVE EYEWEAR AT THE TIME OF OCCURRENCE. THE MATERIAL SAFETY DATA SHEET (MSDS) WAS NOT REVIEWED. THE LABORATORY'S EXPOSURE CONTROL/RISK MANAGEMENT PLANS ARE IN PLACE. THERE WAS NO BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR OPEN WOUNDS. MEDICAL ATTENTION WAS NOT SOUGHT. NO ERRONEOUS RESULTS WERE GENERATED IN CONNECTION TO THIS EVENT.