FDA Adverse Event Malfunction Summary report: N

ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT

MDR report key: 3372231 · Received September 27, 2013

Report

Report Number
9611451-2013-00741
Event Type
Malfunction
Date Received
September 27, 2013
Date of Event
August 30, 2013
Report Date
August 30, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BZE
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE COMPLAINT DEVICE IS CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE IN (B)(4) LAND FOR EVALUATION. WE WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT (B)(4) ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR EVALUATION. IT WAS VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAK. RESULTS: VISUAL INSPECTION REVEALED NO DAMAGE TO THE RETURNED RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT. THE PRESSURE TEST SHOWED THAT THE RETURNED CIRCUIT WAS WITHIN SPECIFICATION. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 130402. CONCLUSION: NO FAULT WAS FOUND WITH THE RETURNED RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT. WE WERE UNABLE TO DETERMINE WHAT MAY HAVE CAUSED THE PROBLEM REPORTED BY THE CUSTOMER. ALL RT340 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS BEFORE BEING RELEASED FOR DISTRIBUTION. ANY BREATHING CIRCUIT WHICH FAILS ANY OF THESE TESTS IS DISCARDED. IN ADDITION, TUBE WEIGHING AND BOND STRENGTH TESTING ARE PERFORMED EVERY 15 MINUTES. IF ANY FAULTS ARE DETECTED THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE VENTILATOR ALARMS". "FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THE EVAQUA EXPIRATORY LIMB WAS LEAKING ON AN RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT. THIS WAS FOUND DURING A VENTILATOR LEAK TEST AND PRIOR TO PATIENT USE.

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THE EVAQUA EXPIRATORY LIMB WAS LEAKING ON AN RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT. THIS WAS FOUND DURING A VENTILATOR LEAK TEST AND PRIOR TO PATIENT USE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
489165 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT BZE BZE FISHER & PAYKEL HEALTHCARE LTD RT340 130402

Patients

Seq Age Sex Outcome Treatment
1