FDA Adverse Event Malfunction Summary report: N

VENTED AUTOFEED CHAMBER

MDR report key: 3372205 · Received September 26, 2013

Report

Report Number
9611451-2013-00734
Event Type
Malfunction
Date Received
September 26, 2013
Date of Event
July 1, 2013
Report Date
August 28, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
BTT
PMA / PMN Number
K934140
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE COMPLAINT MR290V VENTED HUMIDIFICATION CHAMBER WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) WHERE IT WAS VISUALLY INSPECTED FOR THE REPORTED DAMAGE. RESULTS: THE COMPLAINT MR290V CHAMBER WAS RETURNED WITH A BROKEN BAG SPIKE. THE TIP OF THE SPIKE WAS BROKEN. THE SURFACE OF THE BREAK WAS ROUGH (NOT SMOOTHLY CUT). A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 120412. CONCLUSION: BASED ON THE INSPECTION OF THE RETURNED MR290V CHAMBER, THE BAG SPIKE WAS MOST LIKELY SUBJECTED TO EXCESSIVE FORCE. ALL CHAMBERS ARE PRESSURE TESTED BEFORE AND VISUALLY INSPECTED BEFORE THEY LEAVE THE PRODUCTION LINE, ANY BROKEN BAG SPIKES WOULD BE IDENTIFIED DURING THIS PROCESS. ANY CHAMBER THAT FAILS IS REJECTED. THIS SUGGESTS THAT THE DAMAGE TO THE BAG SPIKE OCCURRED AFTER THE CHAMBER WAS RELEASED FOR DISTRIBUTION. THE USER INSTRUCTIONS THAT ACCOMPANY THE MR290 STATE THE FOLLOWING: -"SET APPROPRIATE VENTILATOR ALARMS." -"PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT WATER WAS LEAKING FROM THE CONNECTION PART OF THE WATER FEEDSET TUBE AND THE BAG SPIKE ON AN MR290V VENTED AUTOFEED HUMIDIFICATION CHAMBER. THIS WAS FOUND AFTER A FEW DAYS OF USE. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
486608 VENTED AUTOFEED CHAMBER AUTOFEED CHAMBER BTT FISHER & PAYKEL HEALTHCARE LTD MR290V 1204120304

Patients

Seq Age Sex Outcome Treatment
1 RT200 ADULT DUAL-HEATED BREATHING CIRCUIT