ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2013-00709
- Event Type
- Malfunction
- Date Received
- September 21, 2013
- Date of Event
- August 19, 2013
- Report Date
- August 23, 2013
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE COMPLAINT RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT WAS RETURNED TO FISHER & PAYKEL HEALTHCARE IN (B)(4) WHERE IT WAS VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED A HOLE IN THE EVAQUA EXPIRATORY LIMB APPROXIMATELY 27 CM FROM THE PROXIMAL CONNECTOR ON THE RETURNED RT340 BREATHING CIRCUIT. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT 130403. CONCLUSION: BASED ON THE INSPECTION CONDUCTED, THE EVAQUA EXPIRATORY LIMB WAS MOST LIKELY PUNCTURED BY A BLUNT OBJECT. ALL RT340 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS BEFORE RELEASING FOR DISTRIBUTION. IN ADDITION, TUBE WEIGHING AND BOND STRENGTH TESTING IS PERFORMED EVERY 15 MINUTES. ANY BREATHING CIRCUIT WHICH FAILS ANY OF THESE TESTS IS DISCARDED AND THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. THIS SUGGESTS THAT THE SUBJECT BREATHING CIRCUIT WAS DAMAGED AFTER IT WAS RELEASED FOR DISTRIBUTION. (B)(4). THE USER INSTRUCTIONS SUPPLIED WITH THE RT340 BREATHING CIRCUIT STATE: - "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." - "SET APPROPRIATE VENTILATOR ALARMS". - "FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE." THE HOSPITAL CORRECTLY FOLLOWED OUR USER INSTRUCTIONS AND DETECTED THE REPORTED DAMAGE DURING THE SETUP CHECK AND BEFORE USE ON A PATIENT.
A HEALTHCARE FACILITY IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT A HOLE WAS FOUND ON THE EXPIRATORY LIMB OF AN RT340 ADULT DUAL-HEATED EVAQUA BREATHING CIRCUIT. THIS WAS FOUND DURING A VENTILATOR LEAK TEST AND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 478630 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT340 | 130403 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |