FDA Adverse Event Malfunction Summary report: N

BOND ORACLE HER2 IHC SYSTEM

MDR report key: 3350421 · Received August 27, 2013

Report

Report Number
3004859032-2013-00001
Event Type
Malfunction
Date Received
August 27, 2013
Date of Event
July 29, 2013
Report Date
August 2, 2013
Manufacturer
LEICA BIOSYSTEMS NEWCASTLE LTD.
Product Code
MVC
PMA / PMN Number
PA090015
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE KIT WAS RETURNED TO LEICA BIOSYSTEMS (B)(4) AND INTERNAL TESTING HAS BEEN CARRIED OUT. THE RETURNED CUSTOMER KIT WAS TESTED AGAINST A RETAINED UNIT OF THE SAME LOT AND ALSO AGAINST THE LOT WHICH THE CUSTOMER IS NOW USING. ALTHOUGH RESULTS FOR LOT 20500 SHOW A DIFFERENCE IN STAINING BETWEEN THE RETAINED AND RETURNED KITS WITH REGARDS TO CELL LINES (WITHIN ACCEPTABLE LIMITS) THE IN HOUSE CONTROL AND TEST SLIDES ARE COMPARABLE. DESPITE THE SLIGHT DIFFERENCE IN CONTROL INTENSITY, THE TEST SECTION CONSISTENTLY SCORES 1+ AND THE OUTCOME FOR THE PT WOULD NOT CHANGE OVER THE 3 KITS TESTED. THEREFORE, THERE WOULD BE NO RISK OF POTENTIAL INJURY TO THE PT USING THE RESULTS OBSERVED BY THE CUSTOMER.

Description of Event or Problem · 1

THE NEW ORACLE KIT (LOT 20500) WAS STARTED AND THE LAB STAFF FELT THE STAINING ON THE CELL LINE WAS A LITTLE WEAKER THAN NORMAL, BUT ACCEPTABLE, AND THEY CONTINUED TO USE THE KIT. THE FIRST TWO RUNS THEY WERE HAPPY WITH. THE NEXT TWO RUNS WERE REJECTED AS THE CELL LINES WERE STAINING VERY WEAKLY. THE FIRST REJECTED RUN WAS STAINED ON FREDDIE (B)(4). THE CONTROL SLIDE (B)(4) SHOWED SOME VERY WEAK MEMBRANE STAINING IN THE 1+ AND 2+ WAS ALSO WEAK WITH LESS CELLS STAINING THAN EXPECTED. THE IN HOUSE CONTROL WAS OK. ON THE SECOND REJECTED RUN STRAINING ON GEOFFREY (B)(4), THE CONTROL SLIDE (B)(4) STAINING WAS EVEN WEAKER. THE 1 + WAS NEGATIVE FOR MEMBRANE STAINING WITH VERY WEAK BRUSH BORDERS. THE 2+ AND 3+ WERE ALSO VERY WEAKLY STAINED. THE IN HOUSE CONTROL WAS WEAK BUT STILL GAVE 1+, 2+ AND 3+. AT THIS POINT, THE SITE CONTACTED THE LEICA FIELD SUPPORT SCIENTIST AND ALSO OPENED A NEW KIT TO RUN AS A COMPARISON (RUN ON FREDDIE). THIS KIT (LOT 201437) STAINED FINE (CONTROLS (B)(4) STAINED AS EXPECTED). THE CASES FROM THE REJECTED RUNS HAVE BEEN REPEATED WITH THIS NEW KIT. AS A RESULT OF THE OBSERVED STAINING BY THE CUSTOMER, THERE WAS A TWO DAY DELAY IN DIAGNOSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
420553 BOND ORACLE HER2 IHC SYSTEM NONE MVC LEICA BIOSYSTEMS NEWCASTLE LTD. TA9145 20500

Patients

Seq Age Sex Outcome Treatment
1 Other