FDA Adverse Event Injury Summary report: N

TI MATRIX LOCKING CAP

MDR report key: 3345749 · Received September 12, 2013

Report

Report Number
2530088-2013-01150
Event Type
Injury
Date Received
September 12, 2013
Report Date
August 14, 2013
Manufacturer
SYNTHES BRANDYWINE
Product Code
NKB
PMA / PMN Number
K100952
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ONGOING; NO CONCLUSION COULD BE DRAWN. WITHOUT A LOT NUMBER THE DEVICE HISTORY RECORDS REVIEW COULD NOT BE COMPLETED.

Additional Manufacturer Narrative · 1

DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS MANUFACTURING EVALUATION WAS PERFORMED AND STATES THAT THE IMPLANT WAS RECEIVED ASSEMBLED WITH EXTENSIVE DAMAGE. THERE IS NO FLATTENING OF TEETH ON THE SADDLE INDICATING THE ROD WAS NOT LOOSE ON THIS LOCKING CAP. RAW MATERIAL CANNOT BE MEASURED BECAUSE OF LACK OF SURFACE AREA, BUT RAW MATERIAL WAS VERIFIED AT DHR REVIEW. ALL MEASUREMENTS THAT COULD BE TAKEN WERE FOUND TO MEET SPECIFICATIONS. ALSO, REVIEW OF THE DEVICE HISTORY RECORD REVEALED NO COMPLAINT RELATED ANOMALIES. PLACEHOLDER.

Additional Manufacturer Narrative · 1

ADDITIONAL NARRATIVE: A PRODUCT DEVELOPMENT EVENT EVALUATION WAS CONDUCTED AND IT STATES THAT THE FOUR TITANIUM MATRIX LOCKING CAPS (PART 04.632.000) WERE RETURNED WITH THE CONSTRUCT. THE LOCKING CAPS 04.632.000 ARE FOUND IN THE MATRIX SPINE SYSTEM ((B)(4)). THE CAPS INITIALLY LOCK TO THE POLYAXIAL HEAD TO PREVENT THE ROD FROM FULLY RELEASING FROM THE SCREW HEAD WHILE ALLOWING IT TO SLIDE FREELY. WHEN THE CONSTRUCT IS COMPLETE, THE LOCKING CAP IS TIGHTENED WITH A 10NM (03.620.061) IN ORDER FOR THE RODS TO BE FULLY SECURED FOR A RIGID CONSTRUCT. THERE IS SEVERAL LOT NUMBERS ASSOCIATED WITH THE RETURNED LOCKING CAPS (LOT 7003294, 7035943, 7037094, AND 7035940). THE TITANIUM MATRIX LOCKING CAPS WERE RECEIVED ASSEMBLED. THERE ARE NICKS AND SCRATCHES ON THE SD25 (STARDRIVE) FACE DUE TO TOOL ENGAGEMENT. THERE IS NO EVIDENCE OF CROSS THREADING AND THERE IS NO FLATTENING OF THE TEETH ON THE SADDLE INDICATING THE ROD WAS NOT LOOSE ON THE LOCKING CAPS. THE LOCKING CAPS WERE ENGAGED WITH POLYAXIAL SCREW (04.606.665 LOT 6553706) WITH NO DIFFICULTIES OCCURRING DURING ASSEMBLY TO SCREW. A REVIEW OF THE MOST CURRENT EDITION OF THE DESIGN DRAWING WAS PERFORMED AND THERE WERE NO CHANGES MADE TO THE TITANIUM MATRIX LOCKING CAP ASSEMBLY. THE DESIGN IS ADEQUATE FOR ITS INTENDED USE AND DID NOT CONTRIBUTE TO THIS COMPLAINT CONDITION. THE RETURNED DEVICE DID NOT DEMONSTRATE THE COMPLAINT CONDITION; THEREFORE, THIS COMPLAINT IS INVALID FROM A DESIGN PERSPECTIVE.

Description of Event or Problem · 1

IT WAS REPORTED THAT PATIENT WAS IMPLANTED ON (B)(6) 2012 WITH MATRIX CONSTRUCT AT POSITIONS: LUMBAR 4 TO SACRUM 1 AND ILEUM ON BOTH SIDES. PATIENT WAS RETURNED TO THE O.R. ON (B)(6) 2013 FOR REMOVAL OF HARDWARE. X-RAYS CONFIRMED BOTH OF THE LOCKING CAPS AT THE L3 POSITION BACKED OUT. DURING SURGERY IT WAS OBSERVED THAT THE LOCKING CAPS AT THE L4 POSITION WERE LOOSE. PATIENT WAS REVISED TO COMPETITOR¿S HARDWARE. PROCEDURE WAS COMPLETED WITH NO PROBLEMS. THIS IS 1 OF 10 REPORTS FOR COMPLAINT (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
458598 TI MATRIX LOCKING CAP NKB SYNTHES BRANDYWINE 7003294

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention