FDA Adverse Event
Malfunction
Summary report: N
O.R. CAMERA COVER ADAPTOR FOR BERCHTOLD NON-STERILE
MDR report key: 3345710
·
Received August 8, 2013
Report
- Report Number
- 1836161-2013-00135
- Event Type
- Malfunction
- Date Received
- August 8, 2013
- Date of Event
- July 31, 2013
- Report Date
- August 8, 2013
- Manufacturer
- ASPEN SURGICAL PRODUCTS, INC.
- Product Code
- FTA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
Description of Event or Problem · 1
CUSTOMER REMOVED ADAPTER TO WORK ON THE CAMERA. THE ASPEN ADAPTER HAS 3 FIXED SCREWS THAT SITS IN A GROVE ON THE HOUSING OF THE BERCHTOLD CAMERA. THE ENTIRE CAMERA, INCLUDING THE GROOVE, IS BLACK. HOWEVER WHEN THE ENGINEERING REMOVED THE ADAPTER AND THE CAMERA, THE GROVE WAS ENTIRELY SHAVED DOWN TO THE METAL FROM THE ADAPTER CONTINUOUSLY BEING TURNED BY STAFF MEMBERS. THE ENGINEER IS CONCERNED THAT THE BLACK SHAVING FROM THE COATING OF THE CAMERA IS COMING LOOSE AND COULD POSSIBLY CONTAMINATE THE FIELD/FALL INTO A PT DURING A PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 375114 | O.R. CAMERA COVER ADAPTOR FOR BERCHTOLD NON-STERILE | LIGHT HANDLE | FTA | ASPEN SURGICAL PRODUCTS, INC. | LT-9902 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |