FDA Adverse Event Malfunction Summary report: N

OXYGEN REGULATOR

MDR report key: 33428 · Received June 18, 1996

Report

Report Number
MW4001352
Event Type
Malfunction
Date Received
June 18, 1996
Date of Event
June 15, 1996
Report Date
June 18, 1996
Manufacturer
LIFE SUPPORT PRODUCTS, INC.
Product Code
CAN
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
TX, US
Reporter Occupation
RISK MANAGER

Narratives

Description of Event or Problem · 1

FLASH FIRE FROM CLASS "D" OXYGEN CYLINDER WITH A SUSPECT REGULATOR. FIRE FIGHTER OPENED VALVE ON OXYGEN CYLINDER TO ADMINISTER OXYGEN TO PT DURING AN EMS CALL. FLASH FIRE IMMEDIATELY RESULTED CAUSING SUBSTANTIAL BURNS TO THE ARMS, CHEST, NECK AND FACE OF FIRE FIGHTER. NICKEL SIZE HOLE IN REGULATOR FOUND AFTER FLASH FIRE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 OXYGEN REGULATOR OXYGEN REGULATOR CAN LIFE SUPPORT PRODUCTS, INC. 270

Patients

Seq Age Sex Outcome Treatment
1 NO INFO Hospitalization