FDA Adverse Event Malfunction Summary report: N

ZIMMER SKIN GRAFT MESHER

MDR report key: 3338896 · Received August 22, 2013

Report

Report Number
1526350-2013-00467
Event Type
Malfunction
Date Received
August 22, 2013
Date of Event
July 1, 2013
Report Date
July 26, 2013
Manufacturer
ZIMMER SURGICAL
Product Code
FZW
Removal / Correction Number
RES 61798
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

BEGINNING MAY 31, 2012, US AND CANADIAN CUSTOMERS WERE SENT AN URGENT PATIENT SAFETY ADVISORY INFORMING THEM OF THE NEED FOR PROPER CARE AND PREVENTIVE MAINTENANCE OF THEIR ZIMMER SKIN GRAFT MESHER. CUSTOMERS WERE INFORMED THAT IMPROPERLY MAINTAINED INSTRUMENTS MAY CAUSE DONOR SITE INJURIES OR RESULT IN DAMAGE TO THE GRAFT. CUSTOMERS WERE REQUESTED TO CONTACT ZIMMER TO SCHEDULE MAINTENANCE FOR THE DEVICE IN ACCORDANCE WITH THE INSTRUCTIONS FOR USE. THE DEVICE WAS RETURNED TO THE MANUFACTURER FOR REPAIR AND EVALUATION. THE SERVICE RECORD INDICATES THAT THE DEVICE WAS MANUFACTURED ON 09/20/1993 AND WAS LAST REPAIRED ON 07/02/2001 FOR A BENT COMB. EVALUATION OF THE DEVICE VERIFIED THE COMB TO BE DAMAGED. ADDITIONALLY, THE RIGHT END OF THE ROLLER WAS GNARLED AND GALLED. WEAR TO THE SIDE PLATES WA ALSO OBSERVED. PRIOR TO REPAIR, A TEST MESH AND CALIBRATION CHECK WAS NOT PERFORMED DUE TO THE DAMAGED COMB. THE CAUSE IS MOST LIKELY DUE TO THE USER NOT MAINTAINING THE DEVICE PER PREVENTIVE MAINTENANCE AND HANDLING ACCORDING TO THE INSTRUCTIONS FOR USE. THE ZIMMER SKIN GRAFT MESHER SHOULD BE RETURNED EVERY 12 MONTHS FOR INSPECTION AND PREVENTIVE MAINTENANCE. ANNUAL FACTORY CALIBRATION CHECKS ARE STRONGLY RECOMMENDED TO VERIFY CONTINUED ACCURACY. THE DEVICE WAS SERVICED AND RETURNED TO THE CUSTOMER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE ZIMMER SKIN GRAFT MESHER HAD A BENT GUARD. ADDITIONAL CLINICAL FOLLOW UP WITH THE CUSTOMER INDICATED THAT THE REPORTED ISSUE WAS OBSERVED IN STERILE PROCESSING, PRIOR TO PUTTING THE DEVICE TO INVENTORY. THERE WAS NO PATIENT INVOLVEMENT.