FDA Adverse Event Malfunction Summary report: N

SCISSORS INSERT CEV607 6PK DIA 5MM

MDR report key: 3338747 · Received August 29, 2013

Report

Report Number
9680837-2013-00289
Event Type
Malfunction
Date Received
August 29, 2013
Date of Event
October 31, 2012
Report Date
November 8, 2012
Manufacturer
MEDTRONIC XOMED INSTRUMENTATION S.A.S.
Product Code
GEI
PMA / PMN Number
K993655
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FR
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICES WERE EVALUATED AND INDICATED THAT THE BLADES WERE BLUNT. THE PLASTIC SKIRT WAS DAMAGED. THE BLADES WERE BLUNT RESULTING FROM INSTRUMENT USE. SHARPENING WAS NECESSARY. THE BLACK PLASTIC SKIRT WAS PROBABLY DAMAGED AFTER WITHDRAWAL FROM THE TROCAR. (B)(6). NOTE: THIS MDR IS BEING FILED AS A RESULT OF AN INTERNAL REVIEW OF COMPLAINTS FROM MEDTRONIC'S MXI FACILITY IN (B)(4). THIS REVIEW WAS CONDUCTED TO RESOLVE SEVERAL ISSUES DISCOVERED IN THE MXI COMPLAINT HANDLING PROCESS. ISSUES RESOLVED INCLUDE THE MISCLASSIFICATION OF DEVICES RETURNED FOR REPAIR, AS WELL AS GAPS IN REPORTING. MXI CAPA (B)(4) WAS OPENED TO ADDRESS THESE ISSUES. MEDTRONIC'S REFERENCE #: N/A. (B)(4).

Description of Event or Problem · 1

TWO DEVICES (PART#: CEV607, LOT: 120504) WERE RETURNED TO MXI SERVICE AND REPAIR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
428117 SCISSORS INSERT CEV607 6PK DIA 5MM GEI MEDTRONIC XOMED INSTRUMENTATION S.A.S. CEV607 120504

Patients

Seq Age Sex Outcome Treatment
1