FDA Adverse Event Malfunction Summary report: N

LINEAR ACCELERATOR

MDR report key: 33353 · Received May 8, 1996

Report

Report Number
MW4001336
Event Type
Malfunction
Date Received
May 8, 1996
Date of Event
April 15, 1996
Report Date
April 17, 1996
Manufacturer
VARIAN ASSOC, INC.
Product Code
IYE
Product Problem
Yes
Report Source
Voluntary report
Reporter Location
ID, US
Reporter Occupation
PHYSICIST

Narratives

Description of Event or Problem · 1

THERAPIST OBSERVED, VIA IN ROOM MONITOR, A 2MM SHIFT IN COUCH VERTICAL DISTANCE UP WHEN MAKING A LONGITUDINAL ADJUSTMENT. VERTICAL THUMB WHEEL NOT TOUCHED. ELECTRONIC'S TECH: WHEN DROPPED COUCH SKIRT, NOTICED THAT WHENEVER LATERAL, LONGITUDE, OR COUCH ROTATION WERE DRIVEN A SLIGHT MOVEMENT ON VERTICAL DRIVE MOTOR OCCURRED. SET LONGITUDINAL TABLE TOP FULL OUT, SET COUCH VERTICAL ON 110CM ON MONITOR, DROVE COUCH IN AND COUCH RAISED UP, WITHOUT VERTICAL ACTIVATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LINEAR ACCELERATOR LINEAR ACCELERATOR IYE VARIAN ASSOC, INC.

Patients

Seq Age Sex Outcome Treatment
1 NO INFO