Malfunction
Summary report: N
LINEAR ACCELERATOR
MDR report key: 33353
·
Received May 8, 1996
Report
- Report Number
- MW4001336
- Event Type
- Malfunction
- Date Received
- May 8, 1996
- Date of Event
- April 15, 1996
- Report Date
- April 17, 1996
- Manufacturer
- VARIAN ASSOC, INC.
- Product Code
- IYE
- Product Problem
- Yes
- Report Source
- Voluntary report
- Reporter Location
- ID, US
- Reporter Occupation
- PHYSICIST
Narratives
Description of Event or Problem · 1
THERAPIST OBSERVED, VIA IN ROOM MONITOR, A 2MM SHIFT IN COUCH VERTICAL DISTANCE UP WHEN MAKING A LONGITUDINAL ADJUSTMENT. VERTICAL THUMB WHEEL NOT TOUCHED. ELECTRONIC'S TECH: WHEN DROPPED COUCH SKIRT, NOTICED THAT WHENEVER LATERAL, LONGITUDE, OR COUCH ROTATION WERE DRIVEN A SLIGHT MOVEMENT ON VERTICAL DRIVE MOTOR OCCURRED. SET LONGITUDINAL TABLE TOP FULL OUT, SET COUCH VERTICAL ON 110CM ON MONITOR, DROVE COUCH IN AND COUCH RAISED UP, WITHOUT VERTICAL ACTIVATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LINEAR ACCELERATOR | LINEAR ACCELERATOR | IYE | VARIAN ASSOC, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NO INFO |