FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3334784 · Received September 9, 2013

Report

Report Number
2531779-2013-14799
Event Type
Malfunction
Date Received
September 9, 2013
Report Date
August 19, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/19/2013 WITH THE FOLLOWING FINDINGS: THE KEYPAD WAS FOUND TO BE INTACT; THE KEYPAD BUTTON SYMBOLS WERE FOUND TO BE WORN. INVESTIGATION REVEALED CONTAMINATION FOUND UNDER THE DOWN ARROW AND CONTRAST KEY CONTACTS. ALL KEYPAD BUTTONS WERE FOUND TO BE RESPONSIVE. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.

Description of Event or Problem · 1

THE PUMP WAS RETURNED FOR INVESTIGATION. INVESTIGATION REVEALED CONTAMINATION FOUND UNDER THE DOWN ARROW AND CONTRAST KEY CONTACTS. THERE WAS NO REPORTED ADVERSE EVENT ASSOCIATED WITH THIS COMPLAINT. THIS REPORT IS MADE BASED ON RESULTS OF INVESTIGATION COMPLETED ON 08/19/2013.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
449268 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 51 YR