FDA Adverse Event
Death
Summary report: N
MEDICAL RESEARCH LABORATORIES, INC. (MRL)
MDR report key: 333278
·
Received May 15, 2001
Report
- Report Number
- 333278
- Event Type
- Death
- Date Received
- May 15, 2001
- Date of Event
- May 4, 2001
- Report Date
- May 10, 2001
- Manufacturer
- MEDICAL RESEARCH LABORATORIES, INC.
- Product Code
- MKJ
- Adverse Event
- Yes
- Report Source
- User Facility report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
PT FOUND IN LOCKED CAR WHEN FIRST REACHED BY PARAMEDIC CREW AND FOUND TO BE IN VENTRICULAR FIBRILLATION. THE MRL 360SLX ECG MON/DEFIB WAS SET FOR 360 JOULES AND WAS DISCHARGED X2 SUCCESSFULLY BUT WITHOUT CONVERTING THE ARRHYTHMIA. WHEN THE 3RD CONSECUTIVE SHOCK WAS ATTEMPTED THE BATTERY GAVE OUT AND THE "SPARE" BATTERY WAS ALSO DEPLETED. THE PT WAS TRANSPORTED TO THE NEAREST APPROVED COMPREHENSIVE ER FACILITY UNDER CPR AND 100% OXYGEN. SHORTLY AFTER THE FAILED 3RD SHOCK, THE PT SELF-CONVERTED TO ASYSTOLE AND REMAINED UNTIL DECLARED DEAD ON ARRIVAL.
Description of Event or Problem · 1
ADD'L INFO REC'D FROM MFR 11/7/03: H6. A BROKEN LOW-VOLTAGE MALE PIN FROM THE THERAPY CABLE WAS OBSERVED TO BE STUCK INSIDE THE FEMALE PIN OF THE DEVICE THERAPY CONNECTOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 22853 | MEDICAL RESEARCH LABORATORIES, INC. (MRL) | MRL MODEL SXL360 | MKJ | MEDICAL RESEARCH LABORATORIES, INC. | MRL MODEL SXL360 | * |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 37 YR | Death |