PRECISION®
Report
- Report Number
- 3006630150-2013-01980
- Event Type
- Malfunction
- Date Received
- September 5, 2013
- Date of Event
- June 28, 2012
- Report Date
- June 28, 2012
- Manufacturer
- BOSTON SCIENTIFIC NEUROMODULATION
- Product Code
- LGW
- PMA / PMN Number
- 030017
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENTS INVOLVED IN THE EVENT: MODEL #:SC-2218-50, SERIAL #:(B)(4), DESCRIPTION: LINEAR ST LEAD, 50CM.
ADDITIONAL INFORMATION WAS RECEIVED THAT THE PATIENT UNDERWENT AN EXPLANT PROCEDURE. ADDITIONAL SUSPECT MEDICAL DEVICE COMPONENT INVOLVED IN THE EVENT: MODEL #: SC-8120-70, SERIAL #: (B)(4), DESCRIPTION: ARTISAN 2X8 SURGICAL LEAD - 70 CM. THE EXPLANTED DEVICES WERE NOT RETURNED TO BSN AS THEY WERE DISCARDED BY THE MEDICAL FACILITY. IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE DEVICE HISTORY RECORDS WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MED WATCH WILL BE FILED.
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING ELECTRICITY SHOCKS WHEN STIMULATOR IS ON. DATABASE ANALYSIS ALSO REVEALED HIGH IMPEDANCES ON THE LEADS. DEVICE MALFUNCTION WAS SUSPECTED. THE PATIENT WILL UNDERGO AN EXPLANT PROCEDURE.
A REPORT WAS RECEIVED THAT THE PATIENT WAS EXPERIENCING ELECTRICITY SHOCKS WHEN STIMULATOR IS ON. DATABASE ANALYSIS ALSO REVEALED HIGH IMPEDANCES ON THE LEADS. DEVICE MALFUNCTION WAS SUSPECTED. THE PATIENT WILL UNDERGO AN EXPLANT PROCEDURE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 442380 | PRECISION® | SPINAL CORD STIMULATOR | LGW | BOSTON SCIENTIFIC NEUROMODULATION | SC-1110-02 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR |