FDA Adverse Event Malfunction Summary report: N

OT VERIO IQ METER

MDR report key: 3326128 · Received September 5, 2013

Report

Report Number
2939301-2013-02004
Event Type
Malfunction
Date Received
September 5, 2013
Report Date
August 27, 2013
Manufacturer
LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL
Product Code
NBW
PMA / PMN Number
K110637
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PATIENT

Narratives

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED LIFESCAN (B)(4), ALLEGING THAT THE SUBJECT METER READ INACCURATELY COMPARED TO ANOTHER DEVICE. THE REPORTER CLAIMED OBTAINING BLOOD GLUCOSE READINGS OF ¿193, 155, AND 36 MG/DL¿ WITH THE SUBJECT METER AND ¿177, 224, AND 109 MG/DL¿ RESPECTIVELY ON ANOTHER DEVICE, PERFORMED WITHIN 30 MINUTES OF EACH OTHER. BASED ON STATISTICAL METHODOLOGY, THE CALCULATED DIFFERENCE OF THESE GLUCOSE RESULTS EXCEEDS LIFESCAN¿S CRITERIA FOR ACCURACY. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
441206 OT VERIO IQ METER GLUCOSE MONITORING SYS/KIT NBW LIFESCAN EUROPE, A DIVISION OF CILAG GMBH INTL 3264331

Patients

Seq Age Sex Outcome Treatment
1 69 YR