FDA Adverse Event Malfunction Summary report: N

ADULT INSPIRATORY HEATED BREATHING CIRCUIT

MDR report key: 3325818 · Received September 4, 2013

Report

Report Number
9611451-2013-00665
Event Type
Malfunction
Date Received
September 4, 2013
Report Date
August 8, 2013
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
BTT
PMA / PMN Number
SEE H10
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE RT205 ADULT INSPIRATORY HEATED BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. THE COMPLAINT DEVICES ARE CURRENTLY EN ROUTE TO FISHER & PAYKEL HEALTHCARE IN (B)(4) FOR INVESTIGATION. A FOLLOW UP REPORT WILL BE PROVIDED ONCE WE HAVE COMPLETED OUR INVESTIGATION.

Additional Manufacturer Narrative · 1

(B)(4). THE RT205 ADULT INSPIRATORY HEATED BREATHING CIRCUIT IS NOT SOLD IN THE USA BUT IT IS SIMILAR TO A PRODUCT WHICH IS SOLD IN THE USA. THE 510(K) FOR THAT PRODUCT IS K983112. METHOD: FIVE UNSEALED RT205 ADULT BREATHING CIRCUIT KITS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE FOR INSPECTION. THE RETURNED BREATHING CIRCUITS WERE VISUALLY INSPECTED, PRESSURE TESTED, AND IMMERSED IN A WATER BATH TO TEST FOR LEAKS. RESULTS: THE PRESSURE TEST RESULTS WERE OUT OF SPECIFICATION DUE TO EXCESSIVE LEAK. WHEN THE SUBJECT BREATHING CIRCUITS WERE IMMERSED IN THE WATER BATH, THE LEAK WAS OBSERVED AT THE CONNECTION BETWEEN THE LID AND THE BOWL OF THE WATER TRAP. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT NUMBER 130419 . CONCLUSION: THE BREATHING CIRCUIT WATER TRAP CONSISTS OF A BOWL AND A LID, WHICH CAN BE SEPARATED TO ALLOW THE CAREGIVER TO EMPTY THE WATER TRAP. ALL BREATHING CIRCUITS ARE PRESSURE TESTED FOR LEAKS DURING PRODUCTION AND THOSE THAT FAIL ARE REJECTED. THIS SUGGESTS ANY LEAK MUST HAVE DEVELOPED AFTER IT WAS RELEASED FOR DISTRIBUTION, DURING TRANSPORT, STORAGE OR USE, POSSIBLY BY DISTORTION OF THE WATER TRAP WHEN THE BOWL WAS CONNECTED. OUR USER INSTRUCTIONS THAT ACCOMPANY THE RT205 ADULT INSPIRATORY HEATED BREATHING CIRCUIT STATE THE FOLLOWING: "CHECK ALL CONNECTIONS ARE TIGHT BEFORE USE." "PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT." "SET APPROPRIATE ALARMS."

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED THAT THREE RT205 ADULT INSPIRATORY HEATED BREATHING CIRCUITS FAILED THE VENTILATOR LEAK TEST. THIS WAS OBSERVED BEFORE USE ON A PATIENT.

Description of Event or Problem · 1

A HOSPITAL IN THE UK REPORTED THAT THREE RT205 ADULT INSPIRATORY HEATED BREATHING CIRCUITS FAILED THE VENTILATOR LEAK TEST. THIS WAS OBSERVED BEFORE USE ON A PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
438694 ADULT INSPIRATORY HEATED BREATHING CIRCUIT BTT BTT FISHER & PAYKEL HEALTHCARE LIMITED RT205 130419

Patients

Seq Age Sex Outcome Treatment
1