ROD HOLDER
Report
- Report Number
- 1719045-2013-01923
- Event Type
- Malfunction
- Date Received
- August 30, 2013
- Report Date
- August 7, 2013
- Manufacturer
- SYNTHES MONUMENT
- Product Code
- LXH
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SZ
- Reporter Occupation
- OTHER
Narratives
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. DEVICE IS AN INSTRUMENT AND IS NOT IMPLANTED/EXPLANTED. A REVIEW OF THE DEVICE HISTORY RECORDS HAS BEEN REQUESTED. SUBJECT DEVICE HAS BEEN RECEIVED AND IS CURRENTLY IN THE EVALUATION PROCESS. INVESTIGATION IS ON GOING; NO CONCLUSION COULD BE DRAWN.
A MANUFACTURING EVALUATION WAS CONDUCTED. THE REPORT INDICATES THAT THE INSTRUMED MANUFACTURED THE ROD INTRODUCER ASSEMBLY, PART NUMBER 03.616.048, AND LOT NUMBER 6970076. THE MATERIALS OF THE BENT COMPONENT AND BROKEN KNOB WERE CONFIRMED TO BE CORRECT; HOWEVER THE HEAT TREAT VALUE COULD NOT BE VERIFIED DUE TO THE WALL THICKNESS OF THE COMPONENT. THE PARTS OF THIS LOT WERE INSPECTED FOR FUNCTIONAL REQUIREMENTS AT 100% AND FOUND CONFORMING TO ALL REQUIRED SPECIFICATIONS AT SYNTHES INCOMING FINAL INSPECTION. DUE TO DAMAGE, THESE FUNCTIONAL REQUIREMENTS COULD NOT BE VERIFIED DURING EVALUATION. DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS.
DEVICE WAS USED FOR TREATMENT, NOT DIAGNOSIS. REVIEW OF THE DEVICE HISTORY RECORD SHOWED THAT THERE WERE NO ISSUES DURING THE MANUFACTURE OF THE PRODUCT THAT WOULD CONTRIBUTE TO THIS COMPLAINT CONDITION.NOTE: BLANK FIELDS ON THIS FORM INDICATE INFORMATION THAT IS UNKNOWN, UNAVAILABLE OR UNCHANGED.
DEVICE REPORT FROM SYNTHES (B)(6) REPORTS AN EVENT IN (B)(6) AS FOLLOWS: A LOCKING MECHANISM DID NOT LOCK RIGHT. THIS IS REPORT 1 OF 1 FOR COMPLAINT (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 429191 | ROD HOLDER | LXH | SYNTHES MONUMENT | 6970076 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |